Resumo

Direct immersion SPME-GC-MS-MS was used for the analysis of steroids in water at part-per-trillion(ppt) and lower concentrations. The method was validated and extended to real sample analysis. The method were linear from 0.01 to 5 ng/ml with precision less than 10% relative standard deviation for a steroid mixture at 1 ng/ml. Limit of quantitation and limit of detection was found to be 200- 1200 pg/L and 30-200 pg/L respectively and recoveries ranged from 88-103 %. To understand the extraction efficiency of the fiber, a depletion study was performed. The fiber/ sample partition coefficients for the steroids were determined to be 1.0 x 104 to 1.5 x 104 . The extraction was performed without derivatization or the use of an internal standard. SPMEGC-MS-MS effectively demonstrated ultra-trace level detection of steroids in water.

Descrição

Palavras-chave

Steroids using Solid Phase Microextraction, SPME, Gas Chromatography-Mass Spectrometry-Mass Spectrometry, GC-MS-MS, SPMEGC-MS-MS, water analysis