The attenuation factor (FA) it was verified for three different materials: acrylic, polyethylene and aluminum. The radiation beam used was originating from an equipment of dental radiology of 70 kVp. The measure of the radiation dose transmitted by the attenuators was accomplished interposing individual plates, successively, and measuring te transmission to each thickness, that varied from 0,5 mm to 5,0 mm. Fo the aluminum, two groups of different productions were analyzed: a kit of imported plates used in dosimetry, and other material of national production. The difference among values of FA, obtained for both materials, varied being 2,8% to 7,8% for attenuation thickness among 0,5 mm to 2,0 mm.
Resumo (português)
O fator de atenuação (F.A.) foi verificado para três diferentes materiais: acrílico, polietileno e alumínio. O feixe de radiação utilizado foi proveniente de um equipamento de radiologia odontológica de 70 kVp. A medida da dose de radiação transmitida pelos atenuadores foi realizada interpondo placas individuais, sucessivamente, medindo-se a transmissão a cada espessura, que variou de 0,5 mm a 5,0 mm. Para o alumínio, foram analisados dois conjuntos de fabricações distintas, sendo um kit de placas importadas utilizado em dosimetria e outro com material de fabricação nacional. A diferença entre valores do F.A., obtidos para ambos os materiais, variou entre 2,8% a 7,8% para espessuras de atenuação entre 0,5 mm a 2,0 mm.
Descrição
Palavras-chave
Dosimetry, Medical physics, Atenuation factor, X-rays, Dosimetria, Física médica, Fator de atenuação, Raios-X
Idioma
Português
Como citar
Universitas, v. 1, n. 4, p. 167-187, 2011.