Publicação: In vitro evaluation of conventional and locking miniplate/screw systems for the treatment of mandibular angle fractures
Carregando...
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Churchill Livingstone
Tipo
Artigo
Direito de acesso
Acesso restrito
Resumo
This in vitro study evaluated the influence of the type of miniplate and the number of screws installed in the proximal and distal segments on the stability and resistance of Champy's osteosynthesis in mandibular angle fractures. Sixty polyurethane hemimandibles with bone-like consistency were randomly assigned to four groups (n = 15) and sectioned in the mandibular angle region to simulate fracture. The bone segments were fixed by different osteosynthesis methods using 2.0 mm miniplates and 2.0 mm x 6 mm rnonocortical screws. In groups 1 and 2, two conventional (G1) or locking (G2) screws were installed in each bone segment using a conventional (G1) or a locking (02) straight miniplate; in groups 3 and 4, three conventional (03) or locking (04) screws were installed in the proximal segment and four conventional (G3) or locking (04) screws were installed in the distal segment using a conventional (G3) or a locking (G4) seven-hole straight miniplate. The hemimandibles were loaded in compressive strength until a 4 mm displacement occurred between the segments, vertically or horizontally. Locking plate/screw systems provided significantly greater resistance to displacement than conventional ones (p < .01). Locking miniplates offered more resistance than conventional miniplates. Long locking miniplates provided greater stability than short ones.
Descrição
Palavras-chave
internal rigid fixation, mandibular fractures, traumatology, bone plate
Idioma
Inglês
Como citar
International Journal of Oral and Maxillofacial Surgery. Edinburgh: Churchill Livingstone, v. 39, n. 11, p. 1109-1114, 2010.