Biocompatibility Evaluation of 3 Facial Silicone Elastomers
Carregando...
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Lippincott Williams & Wilkins
Tipo
Artigo
Direito de acesso
Acesso restrito
Resumo
The failure of facial prostheses is caused by limitations in the properties of existing materials, especially the biocompatibility. This study aimed to evaluate the biocompatibility of maxillofacial silicones in subcutaneous tissue of rats. Thirty Wistar rats received subcutaneous implants of 3 maxillofacial silicone elastomers (LIM 6050, MDX 4-4210, and industrial Silastic 732 RTV). A histomorphometric evaluation was conducted to analyze the biocompatibility of the implants. Eight areas of 60.11 mm(2) from the surgical pieces were analyzed. Mesenchymal cells, eosinophils, and foreign-body giant cells were counted. Data were submitted to analysis of variance and Tukey test. Initially, all implanted materials exhibited an acceptable tissue inflammatory response, with tissue reactions varying from light to moderate. Afterward, a fibrous capsule around the silicone was observed. The silicones used in the current study presented biocompatibility and can be used for implantation in both medical and dental areas. Their prosthetic indication is conditioned to their physical properties. Solid silicone is easier to adapt and does not suffer apparent modifications inside the tissues.
Descrição
Palavras-chave
Facial silicone, connective tissue, biocompatibility, inflammation
Idioma
Inglês
Citação
Journal of Craniofacial Surgery. Philadelphia: Lippincott Williams & Wilkins, v. 22, n. 3, p. 837-840, 2011.