Different lubricating oil treatments for the determination of Cu, Cr, Fe, Ni, Sb, Pb, and Zn by HR-CS FAAS
Carregando...
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Taylor & Francis Inc
Tipo
Artigo
Direito de acesso
Acesso restrito
Resumo
Microwave-assisted acid decomposition, direct dilution in kerosene, and oil-in-water emulsion were evaluated as lubricating oil pretreatment procedures for Cu, Cr, Fe, Ni, Pb, Sb, and Zn determination by High-Resolution Continuum Source Flame Atomic Absorption Spectrometry (HR-CS FAAS). For wet digestion, results were compared with those obtained by Flame Atomic Absorption Spectrometry (FAAS). The ultrasound probe used in emulsions sonication contaminated samples with Cr, although better results have been observed for the other six elements in this condition. In general, recovery percentages ranging from 81-106%(Cu), 80-107%(Cr), 85-114%(Fe), 82-116%(Ni), 86-117%(Pb), 85-115%(Sb), and 81-114%(Zn) were obtained. The HR-CS FAAS showed to be faster and more sensitive than FAAS.
Descrição
Palavras-chave
HR-CS FAAS, lubricating oil, metals, sample preparation
Idioma
Inglês
Citação
Analytical Letters. Philadelphia: Taylor & Francis Inc, v. 41, n. 9, p. 1555-1570, 2008.