Off line extraction of phenol from human urine sample with isoamyl alcohol and determination by HPLC

Nenhuma Miniatura disponível

Data

1998-01-01

Autores

Menezes, Manoel Lima de [UNESP]
Demarchi, ACCO

Título da Revista

ISSN da Revista

Título de Volume

Editor

Marcel Dekker Inc

Resumo

This method has been developed for extraction and determination of phenol in a urine sample by high performance liquid chromatography.After acid hydrolysis, the free phenol was extracted with isoamyl alcohol solvent, followed by back extraction with 0.5 mol.L-1 sodium hydroxide solution and analyzed by an isocratic HPLC Varian System, equipped with reverse-phase column (MicroPak-C-18). The mobile phase was acetonitrile in 0.01 mol.L-1 hydrochloric acid solution, (20:80 v/v), and at a now-rate of 1.0 mL.min-1. The chromatogram was monitored at 220 nm in room temperature. The identification was based on retention time and the quantification was performed by automatic peak-area determination, corrected for the external standards method.The recovery was higher than 99.5 % for phenol and reproducibility of method was shown to be 2.3% standard deviation and 5.6% coefficient of variance (n=20). The limit detection was 0.05 mgL(-1) and a range of 0.05 to 20.0 mgL(-1) of phenol for linearity.

Descrição

Palavras-chave

Como citar

Journal of Liquid Chromatography & Related Technologies. New York: Marcel Dekker Inc., v. 21, n. 15, p. 2355-2363, 1998.