Determinação de carbaril utilizando testes ELISA (Enzyme-linked immunosorbent assay) e CLAE com detecção por arranjo de diodos

Carregando...
Imagem de Miniatura

Data

2000-08-01

Autores

Toscano, Ilda A. S. [UNESP]
Ribeiro, M. Lúcia [UNESP]
Rocha, Julio C. [UNESP]
Nunes, Gilvanda S.
Barceló, Damiá

Título da Revista

ISSN da Revista

Título de Volume

Editor

Sociedade Brasileira de Química

Resumo

ELISAs have been applied to pesticide residue analysis due to their high sensitivity and selectivity. However, some ELISAs performance may be affected by matrix components. In this work, ELISA for carbaryl in water samples was checked for interference by naturally occurring fulvic acids. The results suggested that the high fulvic acid concentration ( ³30 mg L-1) and acidic pH conditions (pH 4.0) interfere with the signal detection decreasing the method sensitivity. A dilution of the samples and adjust to pH 8.0 are appropriate to minimize the matrix interferences in the ELISA method. Good correlation between ELISA and HPLC-DAD results was observed.

Descrição

Palavras-chave

ELISA, carbaryl, fulvic acids

Como citar

Química Nova. Sociedade Brasileira de Química, v. 23, n. 4, p. 466-471, 2000.