A comparative study of microstrain around three-morse taper implants with machined and plastic copings under axial loading
dc.contributor.author | de Abreu, Celina Wanderley | |
dc.contributor.author | de Vasconcellos, Luis Gustavo Oliveira [UNESP] | |
dc.contributor.author | Balducci, Ivan [UNESP] | |
dc.contributor.author | Nishioka, Renato Sussumu [UNESP] | |
dc.contributor.institution | Universidade Estadual de Campinas (UNICAMP) | |
dc.contributor.institution | Universidade Estadual Paulista (UNESP) | |
dc.date.accessioned | 2022-04-29T07:13:12Z | |
dc.date.available | 2022-04-29T07:13:12Z | |
dc.date.issued | 2013-01-01 | |
dc.description.abstract | Aim: To evaluate the occurrence of microstrain around morse taper implants in straight configuration under axial load in a cast monoblock framework. Methods: Three implants were inserted in a polyurethane block and microunit abutments were installed on the implants with 20 Ncm torque. Plastic and machined copings were adapted on the preset waxing to fabricate the framework (n=5). Four strain gauges were attached on the upper surface of the block and then each framework was tightened on the abutments and a vertical load of 30 kg was applied to five points of the framework. Results: The data obtained in the strain gauge analysis were subjected to twoway ANOVA and Tukey's test (α=0.05). There was statistically significant difference (p=0.0222) for the factor application point and the mean microstrain values were: application point B 402,04μσ, point A 401.21μσ, point E 390.44μσ , point D 341.76μσ and point C 309.19 μσ. Conclusions: There was no microstrain difference between plastic and machined copings during axial loading. Difference in the application point was observed, but remained within bone physiological limits. | en |
dc.description.affiliation | Department of Prosthodontics Piracicaba Dental School University of Campinas | |
dc.description.affiliation | Department of Dental Materials and Prosthodontics São Josè dos Campos Dental School São Paulo State University | |
dc.description.affiliation | Department of Community Dentistry and Pediatric Clinic São Josè dos Campos Dental School São Paulo State University | |
dc.description.affiliationUnesp | Department of Dental Materials and Prosthodontics São Josè dos Campos Dental School São Paulo State University | |
dc.description.affiliationUnesp | Department of Community Dentistry and Pediatric Clinic São Josè dos Campos Dental School São Paulo State University | |
dc.format.extent | 11-15 | |
dc.identifier.citation | Brazilian Journal of Oral Sciences, v. 9, n. 1, p. 11-15, 2013. | |
dc.identifier.issn | 1677-3217 | |
dc.identifier.issn | 1677-3225 | |
dc.identifier.scopus | 2-s2.0-84891485064 | |
dc.identifier.uri | http://hdl.handle.net/11449/227412 | |
dc.language.iso | eng | |
dc.relation.ispartof | Brazilian Journal of Oral Sciences | |
dc.source | Scopus | |
dc.subject | Axial load | |
dc.subject | Dental implant | |
dc.subject | Fixed prosthesis | |
dc.subject | Morse taper | |
dc.subject | Strain gauge | |
dc.title | A comparative study of microstrain around three-morse taper implants with machined and plastic copings under axial loading | en |
dc.type | Artigo | |
dspace.entity.type | Publication | |
unesp.campus | Universidade Estadual Paulista (UNESP), Instituto de Ciência e Tecnologia, São José dos Campos | pt |
unesp.department | Materiais Odontológicos e Prótese - ICT | pt |
unesp.department | Odontologia Social e Clínica Infantil - ICT | pt |