Logo do repositório
 

Occupational radiation exposure in vascular interventional radiology: A complete evaluation of different body regions

dc.contributor.authorBacchim Neto, Fernando Antonio [UNESP]
dc.contributor.authorAlves, Allan Felipe Fattori [UNESP]
dc.contributor.authorMascarenhas, Yvone Maria
dc.contributor.authorNicolucci, Patrícia
dc.contributor.authorPina, Diana Rodrigues de [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionSapra Landauer
dc.contributor.institutionUniversidade de São Paulo (USP)
dc.date.accessioned2018-12-11T17:06:45Z
dc.date.available2018-12-11T17:06:45Z
dc.date.issued2016-08-01
dc.description.abstractPurpose To perform a complete evaluation on radiation doses, received by primary and assistant medical staff, while performing different vascular interventional radiology procedures. Materials and methods We evaluated dose received in different body regions during three categories of vascular procedures: lower limb angiography (Angiography), lower limb percutaneous transluminal angioplasty (Angioplasty) and stent graft placement for abdominal aortic aneurysm treatment (A. A. A. Treatment). We positioned the dosimeters near the eye lens, thyroid, chest, abdomen, hands, and feet of the interventional physicians. Equivalent dose was compared with annual dose limits for workers in order to determine the maximum number of procedures per year that each physician could perform. We assessed 90 procedures. Results We found the highest equivalent doses in the A. A. A. Treatment, in which 90% of the evaluations indicated at least one region receiving more than 1 mSv per procedure. Angioplasty was the only procedural modality that provided statistically different doses for different professionals, which is an important aspect on regards to radiological protection strategies. In comparison with the dose limits, the most critical region in all procedures was the eye lens. Conclusions Since each body region of the interventionist is exposed to different radiation levels, dose distribution measurements are essential for radiological protection strategies. These results indicate that dosimeters placed in abdomen instead of chest may represent more accurately the whole body doses received by the medical staff. Additional dosimeters and a stationary shield for the eye lens are strongly recommended.en
dc.description.affiliationInstituto de Biociências de Botucatu Departamento de Física e Biofísica UNESP — Universidade Estadual Paulista, Distrito de Rubião Junior S/N
dc.description.affiliationSapra Landauer, Rua Cid Silva César, 600
dc.description.affiliationCIDRA – Centro de Instrumentação Dosimetria e Radioproteção Faculdade de Filosofia Ciências e Letras de Ribeirão Preto USP — Universidade de São Paulo, Av. Bandeirantes, 3900 Bairro Monte Alegre, Ribeirão Preto
dc.description.affiliationDepartamento de Doenças Tropicais e Diagnóstico por Imagem Faculdade de Medicina de Botucatu UNESP — Universidade Estadual Paulista, Distrito de Rubião Junior S/N
dc.description.affiliationUnespInstituto de Biociências de Botucatu Departamento de Física e Biofísica UNESP — Universidade Estadual Paulista, Distrito de Rubião Junior S/N
dc.description.affiliationUnespDepartamento de Doenças Tropicais e Diagnóstico por Imagem Faculdade de Medicina de Botucatu UNESP — Universidade Estadual Paulista, Distrito de Rubião Junior S/N
dc.format.extent1019-1024
dc.identifierhttp://dx.doi.org/10.1016/j.ejmp.2016.06.014
dc.identifier.citationPhysica Medica, v. 32, n. 8, p. 1019-1024, 2016.
dc.identifier.doi10.1016/j.ejmp.2016.06.014
dc.identifier.file2-s2.0-84991206833.pdf
dc.identifier.issn1724-191X
dc.identifier.issn1120-1797
dc.identifier.scopus2-s2.0-84991206833
dc.identifier.urihttp://hdl.handle.net/11449/173604
dc.language.isoeng
dc.relation.ispartofPhysica Medica
dc.relation.ispartofsjr0,792
dc.rights.accessRightsAcesso aberto
dc.sourceScopus
dc.subjectInterventional radiology
dc.subjectOccupational exposure
dc.subjectRadiation protection
dc.titleOccupational radiation exposure in vascular interventional radiology: A complete evaluation of different body regionsen
dc.typeArtigo
dspace.entity.typePublication
unesp.campusUniversidade Estadual Paulista (UNESP), Faculdade de Medicina, Botucatupt
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Biociências, Botucatupt
unesp.departmentDoenças Tropicais e Diagnósticos por Imagem - FMBpt
unesp.departmentFísica e Biofísica - IBBpt