Efficacy of sealing agents on preload maintenance of screw-retained implant-supported prostheses
| dc.contributor.author | Seloto, Camila Berbel [UNESP] | |
| dc.contributor.author | Sahyon, Henrico Badaoui Strazzi [UNESP] | |
| dc.contributor.author | dos Santos, Paulo Henrique [UNESP] | |
| dc.contributor.author | Delben, Juliana Aparecida | |
| dc.contributor.author | Assunção, Wirley Gonçalves [UNESP] | |
| dc.contributor.institution | Universidade Estadual Paulista (Unesp) | |
| dc.contributor.institution | Ingá Faculty (UNINGá) | |
| dc.date.accessioned | 2018-12-11T17:17:54Z | |
| dc.date.available | 2018-12-11T17:17:54Z | |
| dc.date.issued | 2018-01-01 | |
| dc.description.abstract | Purpose: The aim of this study was to evaluate the effect of sealing agents on preload maintenance of screw joints. Materials and Methods: A total of four groups (n = 10 in each group) of abutment/implant systems, including external hexagon implants and antirotational UCLA abutments with a metallic collar in cobalt-chromium alloy, were assessed. In the control group (CG), no sealing agent was used at the abutment screw/implant interface. In the other groups, three different sealing agents were used at the abutment screw/implant interface: anaerobic sealing agent for medium torque (ASMT), anaerobic sealing agent for high torque (ASHT), and cyanoacrylate-based bonding agent (CYAB). All abutments were attached to the implants at 32 ± 1 N.cm. After 48 ± 2 hours of initial tightening, loosing torque (detorque) was measured using a digital torque wrench. Data were analyzed using Shapiro-Wilk, Wilcoxon, and Kruskal-Wallis tests, at 5% level of significance. Results: In the CG and ASMT groups, detorque was lower than the insertion torque (24.6 ± 1.5 N.cm and 24.3 ± 1.1 N.cm, respectively). In the ASHT and CYAB groups, mean detorque increased in comparison to the insertion torque (51.0 ± 7.4 N.cm and 47.7 ± 15.1 N.cm, respectively). Conclusion: The ASHT was more efficient than the other sealing agents, increasing the remaining preload (detorque value) 58.88%. Although the cyanoacrylate-based bonding agent also generated high detorque values, the high standard deviation suggested its lower reliability. | en |
| dc.description.affiliation | Department of Dental Materials and Prosthodontics Araçatuba Dental School São Paulo State University (UNESP) | |
| dc.description.affiliation | Department of Dentistry Ingá Faculty (UNINGá) | |
| dc.description.affiliationUnesp | Department of Dental Materials and Prosthodontics Araçatuba Dental School São Paulo State University (UNESP) | |
| dc.description.sponsorship | Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) | |
| dc.description.sponsorshipId | CAPES: 166000/2013-2 | |
| dc.format.extent | 123-126 | |
| dc.identifier | http://dx.doi.org/10.11607/jomi.5576 | |
| dc.identifier.citation | International Journal of Oral and Maxillofacial Implants, v. 33, n. 1, p. 123-126, 2018. | |
| dc.identifier.doi | 10.11607/jomi.5576 | |
| dc.identifier.issn | 0882-2786 | |
| dc.identifier.scopus | 2-s2.0-85041818563 | |
| dc.identifier.uri | http://hdl.handle.net/11449/175863 | |
| dc.language.iso | eng | |
| dc.relation.ispartof | International Journal of Oral and Maxillofacial Implants | |
| dc.relation.ispartofsjr | 1,576 | |
| dc.rights.accessRights | Acesso restrito | pt |
| dc.source | Scopus | |
| dc.subject | Detorque | |
| dc.subject | Implant-supported restorations | |
| dc.subject | Preload | |
| dc.subject | Screw joint | |
| dc.title | Efficacy of sealing agents on preload maintenance of screw-retained implant-supported prostheses | en |
| dc.type | Artigo | pt |
| dspace.entity.type | Publication | |
| relation.isDepartmentOfPublication | 5f53b343-da2a-4737-96ec-6e2389a6d704 | |
| relation.isDepartmentOfPublication.latestForDiscovery | 5f53b343-da2a-4737-96ec-6e2389a6d704 | |
| relation.isOrgUnitOfPublication | 8b3335a4-1163-438a-a0e2-921a46e0380d | |
| relation.isOrgUnitOfPublication.latestForDiscovery | 8b3335a4-1163-438a-a0e2-921a46e0380d | |
| unesp.campus | Universidade Estadual Paulista (UNESP), Faculdade de Odontologia, Araçatuba | pt |
| unesp.department | Materiais odontológicos e Prótese - FOA | pt |