Fabrication and characterization of ferrocenece containing hydrogel for glucose biosensor application

Nenhuma Miniatura disponível

Data

2014-01-01

Autores

Crulhas, Bruno R. [UNESP]
Ramos, Naira P. [UNESP]
Basso, Caroline R. [UNESP]
Costa, Vladimir E. [UNESP]
Castro, Gustavo R. [UNESP]
Pedrosa, Valber A. [UNESP]

Título da Revista

ISSN da Revista

Título de Volume

Editor

Resumo

A detailed electrochemical study of a novel microsystem biosensor based on ferrocene (Fc) intercalated poly(ethylene) glycol (PEG) and its potential use in the development of amperometric biosensors is presented. A redox hydrogel electrode, fabricated by covering the PEG-FC on the surface of gold microelectrode, exhibits a quasi reversible electrochemical characteristic. The characterization and dependence of Fc concentration to construct the biosensor were carried out by cyclic voltammetry measurements. The sensitivity, repeatability and linearity of the new glucose biosensor are found to be excellent over earlier reported glucose biosensor.

Descrição

Palavras-chave

Biosensor, Glucose detection, Redox mediator

Como citar

International Journal of Electrochemical Science, v. 9, n. 12, p. 7596-7604, 2014.