Linear relationship between potential and concentration in flow potentiometry

Nenhuma Miniatura disponível

Data

2000-03-01

Autores

Neto, JAG
Zagatto, EAG

Título da Revista

ISSN da Revista

Título de Volume

Editor

Wiley-Blackwell

Resumo

In potentiometric-flow systems, linear-potential responses for logarithmic concentrations can be attained for first-(or pseudo-first-) order reactions in which the monitored chemical species react with the analyte during a fixed time interval. To demonstrate this property, the determination of glycerol based on its oxidation by periodate and potentiometric monitoring of the remaining periodate was selected. Influence of reagent concentration and timing on the linearity of the analytical curve were investigated. A mathematical treatment was derived, and potentialities/limitations of the approach were outlined. The system was applied to analysis of soap and lixivia samples. The analytical curve within 200 and 2000 mg L-1 (r = 0.99975; n = 5) was described as E = 8.166 + 0.0478 (glycerol). The sample throughput was 100 h(-1), and a measurement repeatability within 0.5 mV was always observed. By applying a t-test, there was no statistical difference between the results obtained by the proposed procedure and by iodimetric titration at the 95% confidence level. (C) 2000 John Wiley & Sons, Inc. Lab Robotics and Automation 12:41-45, 2000.

Descrição

Palavras-chave

Como citar

Laboratory Robotics and Automation. New York: John Wiley & Sons Inc., v. 12, n. 1, p. 41-45, 2000.