Simultaneous Determination of Cadmium and Lead in Medicinal Plants Using Graphite Furnace Atomic Absorption Spectrometry and Direct Slurry Sampling

Nenhuma Miniatura disponível

Data

2012-07-01

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Perkin-elmer Corp

Tipo

Artigo

Direito de acesso

Acesso restrito

Resumo

A method is proposed for the simultaneous determination Cd and Pb in medicinal plants using direct slurry sampling by GF-AAS. The thermal behavior of analytes was checked in aqueous and slurry media using Ir permanent modifier. The optimized pyrolysis (1000 degrees C) and atomization (2200 degrees C) temperatures allowed calibration using aqueous standards.For a 20-mu L injected volume, the calculated characteristic mass was 1.8 pg for Cd and 59 pg for Pb. The relative standard deviation (n=12) for Cd and Pb was 5.3% and 3.5%, respectively. The limit of detection was 0.12 mu g 1,1 for Cd and 1.22 mu g L-1 for Pb. Accuracy was checked by analyzing the standard reference materials Ginkgo Biloba Leaves, Peach Leaves, and Pine Needles. The results obtained for the analytes were in agreement with the certified values at the 95% confidence level. Ten medicinal plant samples were analyzed and the found concentrations ranged from 0.19-0.34 mu g g(-1) for Cd and 1.02-1.91 mu g g(-1) for Pb.

Descrição

Palavras-chave

Idioma

Inglês

Como citar

Atomic Spectroscopy. Shelton: Perkin-elmer Corp, v. 33, n. 4, p. 138-142, 2012.