Silveira, M. M. [UNESP]Fonseca, Luiz Marcos da [UNESP]2014-05-272014-05-272003-12-01Revista de Ciencias Farmaceuticas, v. 24, n. 1, p. 71-78, 2003.0101-3793http://hdl.handle.net/11449/67583The Abbott Cell-Dyn 3000 automated hematological analyzer prepares a histogram of platelet volume, which is measured by a technique using electrical impedance, inferring from its platelet count (PLT), mean platelet volume (MPV) and platelet distribution width (PDW). This equipment also calculates the plateletcrit (PCT). These platelet parameters may be important to evaluate platelet function, but they require standardization, because platelets swell when in contact with ethylenediaminetetra-acetic acid salts, hence an increase in blood sample storage time produces artificially increased results. To assess the effect of storage time on MPV, PLT, PDW and PCT, blood samples from 23 sickle cell anemia patients during steady state (Group I) and 50 from healthy controls (Group II) were placed in Vacutainer® tubes with dipotassium ethylenediaminetetra-acetic acid and measured over a period of 1440 minutes (24 h) at the following times: immediately after the venipuncture (time 0), 15, 30, 60, 120, 240, 360, 480 and 1440 minutes. The mean values of MPV and PCT were significantly increased (p<0.00001) along the storage time in both groups. The mean values of PLT and PDW were practically stable (p>0.05) throughout the storage time in both groups.71-78engBlood sample storage timeEthylenediaminetetra- acetic acidMean platelet volumePlateletcritStandardizationblood analysisblood samplingblood storagecell swellingcontrolled studydiagnostic valuehistogramhumanhuman cellimpedancemean platelet volumeplatelet distribution widthsickle cell anemiastandardizationstorage timethrombocytethrombocyte countthrombocyte preservationthrombocyte volumetimevein punctureStandardization of platelet parameters determination in blood samples collected with EDTAArtigoAcesso restrito2-s2.0-46442566404419635633356792