Linearity and carryover of automated hematology counter Horiba® Abx Micros ESV 60

dc.contributor.authorCruz, Nathan R.N. [UNESP]
dc.contributor.authorYamazaki, Matheus A. [UNESP]
dc.contributor.authorMartinelli, Amanda E. [UNESP]
dc.contributor.authorMartinato, Fernanda [UNESP]
dc.contributor.authorSantana, Aureo E. [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2018-12-11T17:20:07Z
dc.date.available2018-12-11T17:20:07Z
dc.date.issued2018-02-01
dc.description.abstractThe automated hematology counter ABX Micros 60 ESX (Horiba Medical 2012) is veterinary hematology analyzer multi-species that carries 50 counts per hour releases 18 blood parameters, in addition to graphical representations (histograms) for leukocytes, erythrocytes and platelets. The objective of this study is to evaluate the performance of the apparatus with respect to linearity and carryover through commercial control and blood samples from the routine of Veterinary Clinical Pathology Laboratory. According to results we can say that this device has excellent linearity coefficient (r2=0.99) in leukocyte parameters, erythrocytes and platelets during that time dilutions. Regarding the carryover was excellent device performance, however, was not in conformity values in the parameters of MCHC and VPM in one of the methodologies made that can be justified by the limited formula that does not consider the equipment characteristics.en
dc.description.affiliationLaboratório de Patologia Clínica Veterinária Hospital Veterinário Governador 'Laudo Natel' Universidade Estadual Paulista (Unesp) Faculdade de Ciências Agrárias e Veterinárias (FCAV), Via de Acesso Rodovia Paulo Donato Castellane s/n Km 5
dc.description.affiliationUnespLaboratório de Patologia Clínica Veterinária Hospital Veterinário Governador 'Laudo Natel' Universidade Estadual Paulista (Unesp) Faculdade de Ciências Agrárias e Veterinárias (FCAV), Via de Acesso Rodovia Paulo Donato Castellane s/n Km 5
dc.format.extent345-349
dc.identifierhttp://dx.doi.org/10.1590/1678-5150-PVB-5032
dc.identifier.citationPesquisa Veterinaria Brasileira, v. 38, n. 2, p. 345-349, 2018.
dc.identifier.doi10.1590/1678-5150-PVB-5032
dc.identifier.fileS0100-736X2018000200345.pdf
dc.identifier.issn1678-5150
dc.identifier.issn0100-736X
dc.identifier.scieloS0100-736X2018000200345
dc.identifier.scopus2-s2.0-85046864611
dc.identifier.urihttp://hdl.handle.net/11449/176320
dc.language.isopor
dc.relation.ispartofPesquisa Veterinaria Brasileira
dc.rights.accessRightsAcesso aberto
dc.sourceScopus
dc.subjectAutomated hematology
dc.subjectCell blood counter
dc.subjectCounter Horiba® Abx Micros ESV 60
dc.subjectPerformance
dc.subjectVeterinary
dc.titleLinearity and carryover of automated hematology counter Horiba® Abx Micros ESV 60en
dc.titleLinearidade e carryover do contador hematológico Horiba® Abx Micros ESV 60pt
dc.typeArtigo
unesp.author.lattes8905659987198068[5]
unesp.author.orcid0000-0003-4539-1802[5]

Arquivos

Pacote Original
Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
S0100-736X2018000200345.pdf
Tamanho:
845.25 KB
Formato:
Adobe Portable Document Format