Determination of Zinc in Lubricating Oil by Flame MS Employing Ultrasonic Extraction
Loading...
External sources
External sources
Date
Advisor
Coadvisor
Graduate program
Undergraduate course
Journal Title
Journal ISSN
Volume Title
Publisher
Perkin-elmer Corp
Type
Article
Access right
Acesso restrito
External sources
External sources
Abstract
The determination of zinc in lubricating oils by line-source flame atomic absorption spectrometry (LS-FAAS) is proposed employing an alternative sample preparation procedure based on ultrasonic wave-assisted acid extraction. The chemometric optimization of the main parameters influencing the extraction provided the following conditions: 1.3 mL of HCl, 5.0 mL of H(2)O(2), 120 seconds sonication time, 60 degrees C heating temperature, and 45% sonication amplitude of the ultrasound probe. Analytical curves in the 0.5 - 4.0 mg L(-1) Zn range were consistently obtained (R = 0.9990). Commercial unused lubricating oil samples were analyzed by the proposed method and by high-resolution continuum source flame AAS (HR-CS FAAS) as the comparative technique. A paired t-test showed that the results between the two techniques were in agreement at the 95% confidence level. The accuracy of the proposed method was evaluated using an alternative sample preparation method based on recovery tests. Recoveries in the 92 - 119% interval were found. The relative standard deviation was 4% and the limit of detection was 17.6 mu g L(-1).
Description
Keywords
Language
English
Citation
Atomic Spectroscopy. Shelton: Perkin-elmer Corp, v. 32, n. 6, p. 240-245, 2011.