Publicação: Determination of anticonvulsants in human plasma using SPME in a heated interface coupled online to liquid chromatography (SPME-LC)
Carregando...
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Tipo
Artigo
Direito de acesso
Acesso restrito
Resumo
A simple and sensitive method using solid phase microextraction (SPME) and liquid chromatography (LC) with heated online desorption (SPME-LC) was developed and validated to analyze anticonvulsants (AEDs) in human plasma samples. A heated lab-made interface chamber was used in the desorption procedure, which allowed the transference of the whole extracted sample. The SPME conditions were optimized by applying an experimental design. Important factors are discussed such as fiber coating types, pH, extraction time and desorption conditions. The drugs were analyzed by LC, using a C18 column (150 mm 4.6 mm 5 mm); and 50 mmol L1 , pH ¼ 5.50 ammonium acetate buffer : acetonitrile : methanol (55 : 22 : 23 v/v) as the mobile phase with a flow rate of 0.8 mL min1 . The suggested method presented precision (intra-assay and inter-assay), linearity and limit of quantification (LOQ) all adequate for the therapeutic drug monitoring (TDM) of AEDs in plasma.
Descrição
Palavras-chave
Idioma
Inglês
Como citar
Analytical Methods, v. 4, p. 1519-1524, 2012.