Publicação: Amperometric Immunosensor for Chagas' Disease Using Gold CD-R Transducer
Carregando...
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Wiley-Blackwell
Tipo
Artigo
Direito de acesso
Acesso restrito
Resumo
Chagas' disease has been spread mainly through blood transfusion or blood components, therefore the development of a device for neglected diseases with high specificity and sensitivity is important to public health. This work describes the development of an amperometric immunosensor for determination of Chagas' disease through a gold-based electrode obtained from a recordable compact disc modified with 4-(methylmercapto) benzaldehyde for the immobilization of Tc85 protein of the Trypanosoma cruzi. The immunoassays were carried out using positive and negative sera from Chagas' disease patients and immunoglobulin conjugated with peroxidase enzyme. The immunosensor presented -0.949 mu A as cut-off value and was applied in sera samples.
Descrição
Palavras-chave
Chagas' disease, Gold CDtrodes, Immobilization, Self-assembled monolayer, Sensors
Idioma
Inglês
Como citar
Electroanalysis. Malden: Wiley-blackwell, v. 23, n. 11, p. 2555-2561, 2011.