Repository logo

Reversible intermittent flow-injection determination of mercury in sediments and vinasses by cold vapor atomic absorption spectrometry

Loading...
Thumbnail Image

Advisor

Coadvisor

Graduate program

Undergraduate course

Journal Title

Journal ISSN

Volume Title

Publisher

Type

Article

Access right

Abstract

A reversible intermittent flow-injection procedure is proposed for the automated determination of mercury in sediments and vinasses by cold vapor atomic absorption spectrometry, CVAAS. Solutions of sample and stannous chloride are carried by two air streams and sequentially injected into the generator/separator chamber in a segmented asynchronous merging zone configuration. The intermittent flow in the forward direction carries the mercury vapor through the flow cell, and in the backward direction, it aspirates the remaining solution from the vessel to waste. We investigated composition and concentration of reagents, flow rates, commutation times, reactor configuration, and conditions for mercury release. The accuracy was checked by mercury determination in a certified sediment and spiked vinasses and river waters. The system handles about 100 samples per hour (0.50-5.00 μg L-1), consuming ca. 2.5 mL of sample and 50 mg of SnCl2 per determination. Good recoveries (92-103%) were obtained with spiked samples. Results are precise (RSD <3% for 2.5 μg Hg L-1, n = 12) and in agreement with values for certified reference material at 95% confidence level.

Description

Keywords

Language

English

Citation

Laboratory Robotics and Automation, v. 11, n. 5, p. 304-310, 1999.

Related itens

Sponsors

Units

Item type:Unit,
Instituto de Química
IQAR
Campus: Araraquara


Departments

Undergraduate courses

Graduate programs

Other forms of access