Logotipo do repositório
 

Publicação:
A pH optode based on thymol blue: Application to determination of co2 using flow injection analy sis system

dc.contributor.authorSotomayor, M. D.P.T. [UNESP]
dc.contributor.authorRaimundo, Jr.
dc.contributor.authorRohwedder, J. J.R.
dc.contributor.authorNeto, G. Oliveira
dc.contributor.institutionUniversidade Estadual Paulista (UNESP)
dc.contributor.institutionUniversidade Estadual de Campinas (UNICAMP)
dc.contributor.institutionSecretaria Municipal de Educação
dc.date.accessioned2022-04-28T19:30:00Z
dc.date.available2022-04-28T19:30:00Z
dc.date.issued2010-01-01
dc.description.abstractAn optode based on thymol blue (TB), an acid-based indicator, has been constructed and evaluated as a detector in FIA system for CO2determination. The dye was chemically immobilised on the surface of a bifurcated glass optical fibre bundle, using silanisation in organic media. In FIA system, hydrogen carbonate or carbonate samples are injected in a buffer carrier solution, and then are mixed with phosphoric acid solution to generate CO2, which diffuses through a PTFE membrane, in order to be collected in an acceptor carrier fluid, pumped towards to detection cell, in which the optode was adapted. The proposed system presents two linear response ranges, from 1.0 x 10-3to 1.0 x 10-2mol l-1, and from 2.0 x 10-2to 0.10 mol l-1. The sampling frequency was 11 sample h-1, with good repeatability (R.S.D < 4 %, n = 10). In flow conditions the optode lifetime was 170 h. The system was applied in the analysis of commercial mineral water and the results obtained in the hydrogen carbonate determination did not differ significantly from those obtained by potentiometry, at a confidence level of 95 %.en
dc.description.affiliationInstituto de Química; Unesp, São Paulo
dc.description.affiliationInstituto de Química; Unicamp, São Paulo
dc.description.affiliationSecretaria Municipal de Educação, São Paulo
dc.description.affiliationUnespInstituto de Química; Unesp, São Paulo
dc.format.extent33-43
dc.identifierhttp://dx.doi.org/10.26850/1678-4618EQJ.V35.2.2010.P33-43
dc.identifier.citationEcletica Quimica, v. 35, n. 2, p. 33-43, 2010.
dc.identifier.doi10.26850/1678-4618EQJ.V35.2.2010.P33-43
dc.identifier.issn1678-4618
dc.identifier.issn0100-4670
dc.identifier.scopus2-s2.0-85100990286
dc.identifier.urihttp://hdl.handle.net/11449/221676
dc.language.isoeng
dc.relation.ispartofEcletica Quimica
dc.sourceScopus
dc.subjectCO2
dc.subjectFIA
dc.subjectpH-optode
dc.subjectSilanisation
dc.subjectThymol blue
dc.titleA pH optode based on thymol blue: Application to determination of co2 using flow injection analy sis systemen
dc.typeArtigopt
dspace.entity.typePublication
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Química, Araraquarapt

Arquivos