Electrical modelling and power quality analysis of three-phase X-ray apparatus
Carregando...
Arquivos
Fontes externas
Fontes externas
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Tipo
Trabalho apresentado em evento
Direito de acesso
Acesso aberto

Arquivos
Fontes externas
Fontes externas
Resumo
This article proposes the use of PSCAD/EMTDC to model and simulate a three-phase x-ray machine. The model is an alternative tool to power quality studies, load connection impact analysis and power factor correction in the electric power distribution systems. The developed model is used to identify and quantify potential impacts of the X-ray machine in the electric power distribution system. Analyzes and discussions are based on actual measurements obtained in an X-ray apparatus terminals. © 2014 IEEE.
Descrição
Palavras-chave
harmonics, Modelling, Non-linear loads, Power Quality, X-ray
Idioma
Inglês
Citação