Logotipo do repositório
 

Publicação:
Use of FIA systems for on-line dilution in multielement determination by ICP-MS

dc.contributor.authorNeto, José Anchieta Gomes [UNESP]
dc.contributor.authorSilva, José Bento Borba
dc.contributor.authorSouza, Ivan Gonçalves
dc.contributor.authorCurtius, Adilson José
dc.contributor.institutionUniversidade Estadual Paulista (UNESP)
dc.contributor.institutionUniversidade Federal de Santa Catarina (UFSC)
dc.date.accessioned2022-04-28T19:53:28Z
dc.date.available2022-04-28T19:53:28Z
dc.date.issued1999-01-01
dc.description.abstractA flow-injection (FI) system to match concentrations was used as an auto-diluter in multielement determination by inductively coupled plasma-mass spectrometry (ICP-MS). The flow system comprised loop-based injection or a timed valve that introduced a variable sample volume into a spray chamber through a standard Meinhard nebulizer of an ICP-MS. Routinely analyzed samples such as water, plant, and steel were selected. The accuracy of multielement determination was checked against water standard reference material from the National Institute of Standards and Technology (1643d), plant standard reference material from the National Bureau of Standards (1572 citrus leaves), and steel standard reference material from the National Bureau of Standards (AISI 4340). The measuring system was calibrated with a multielement solution, yielding a linear plot with good precision [relative standard deviation (RSD) <3%, n = 12]. The results were in agreement at a 95% confidence level with the certified values for the reference materials and also with those obtained by continuous aspiration and by (FI) with a discrete volume.en
dc.description.affiliationDepto. de Quim. Analítica Instituto de Química UNESP, PO Box 355, 14800-900, Araraquara, SP
dc.description.affiliationDepartamento de Química Univ. Federal de Santa Catarina, PO Box 476, 88040-900, Florianópolis, SC
dc.description.affiliationUnespDepto. de Quim. Analítica Instituto de Química UNESP, PO Box 355, 14800-900, Araraquara, SP
dc.format.extent240-247
dc.identifierhttp://dx.doi.org/10.1002/(SICI)1098-2728(1999)11:4<240
dc.identifier.citationLaboratory Robotics and Automation, v. 11, n. 4, p. 240-247, 1999.
dc.identifier.doi10.1002/(SICI)1098-2728(1999)11:4<240
dc.identifier.issn0895-7533
dc.identifier.scopus2-s2.0-0007915886
dc.identifier.urihttp://hdl.handle.net/11449/223870
dc.language.isoeng
dc.relation.ispartofLaboratory Robotics and Automation
dc.sourceScopus
dc.titleUse of FIA systems for on-line dilution in multielement determination by ICP-MSen
dc.typeArtigopt
dspace.entity.typePublication
relation.isOrgUnitOfPublicationbc74a1ce-4c4c-4dad-8378-83962d76c4fd
relation.isOrgUnitOfPublication.latestForDiscoverybc74a1ce-4c4c-4dad-8378-83962d76c4fd
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Química, Araraquarapt
unesp.departmentQuímica Analítica - IQARpt

Arquivos