Publicação: Braquetes cerâmicos fabricados por impressão 3D: ensaio mecânico de resistência à torção pela aplicação de torques
dc.contributor.advisor | Balducci, Ivan [Unesp] | |
dc.contributor.author | Silva, Paulo César da | |
dc.contributor.institution | Universidade Estadual Paulista (Unesp) | |
dc.date.accessioned | 2022-10-04T16:35:11Z | |
dc.date.available | 2022-10-04T16:35:11Z | |
dc.date.issued | 2022-08-24 | |
dc.description.abstract | A proposição deste estudo foi comparar a nova geração de braquetes personalizados cerâmicos impressos 3D Lightforce Clear , com os braquetes cerâmicos injetados Clarity e GC Chic em um ensaio mecânico de resistência à torção pela aplicação de torque. Amostras de braquetes cerâmicos impressos de alumina policristalina de incisivos centrais superiores (n=10), o Lightforce Clear (Lightforce Orthodontics), e os braquetes cerâmicos injetados de alumina policristalina, Clarity (3M Unitek) e GC Chic (GC Orthodontics) os quais foram submetidos ao teste de resistência à fratura torcional utilizando um segmento de fio de aço inoxidável 0,019 x 0,025 polegadas em uma máquina de ensaio universal (EMIC, DL-1000) à uma velocidade de 1mm/min e célula de carga de 1000 kg. Os resultados foram submetidos à análise estatística ANOVA fator 1. Os valores médios e desvio padrão obtidos no ensaio de resistência à torção de carga máxima de fratura (N) nos três grupos de bráquetes foram: 14,53 ± 2,85 para Lightforce Clear; 12,03 ± 2,44 para Clarity e 6,65 ± 0,35 para o CG Chic. Foi observado que houve diferença estatística significante entre os grupos testados (p<0,05). Os braquetes Lightforce Clear cerâmicos impressos 3D customizados mostraram-se mais resistentes que os injetados apontando como uma alternativa factível e promissora em busca de uma ortodontia digital de precisão contudo existe a necessidade de incorporar novas pesquisas para o avanço desta tecnologia. | pt |
dc.description.abstract | The purpose of this study was to compare the new generation of customized ceramic brackets 3D printed Lightforce Clear, with the injected ceramic brackets Clarity and GC Chic in a mechanical test of torsion resistance by the application of torque. Samples of ceramic printed polycrystalline alumina brackets of upper central incisors (n=10), Lightforce Clear (Lightforce Orthodontics), and the injected ceramic brackets made from polycrystalline alumina, Clarity (3M Unitek) and GC Chic (GC Orthodontics) which were subjected to torsional fracture strength testing using a 0.019 x 0.025-inch stainless steel wire segment in a universal testing machine (EMIC, DL-1000) at a speed of 1 mm/min and a load cell of 1000 kg. The results were submitted to ANOVA factor 1 statistical analysis. The mean values and standard deviation obtained in the maximum fracture load (N) torsional strength test in the three groups of brackets were: 14.53 ± 2.85 for Lightforce Clear; 12.03 ± 2.44 for Clarity and 6.65 ± 0.35 for CG Chic. It was observed that there was a statistically significant difference between the groups tested (p<0.05). The customized 3D printed Lightforce Clear ceramic brackets proved to be more resistant than the injected ones, pointing as a feasible and promising alternative in the search for precision digital orthodontics, however there is a need to incorporate new research to advance this technology. | en |
dc.identifier.capes | 33004145082P6 | |
dc.identifier.uri | http://hdl.handle.net/11449/236811 | |
dc.language.iso | por | |
dc.publisher | Universidade Estadual Paulista (Unesp) | |
dc.rights.accessRights | Acesso aberto | |
dc.subject | braquetes cerâmicos | pt |
dc.subject | resistência à torção | pt |
dc.subject | impressão 3D | pt |
dc.subject | torque | pt |
dc.subject | ceramic brackets | en |
dc.subject | torsion resistance | en |
dc.subject | 3D printing | en |
dc.title | Braquetes cerâmicos fabricados por impressão 3D: ensaio mecânico de resistência à torção pela aplicação de torques | pt |
dc.title.alternative | Ceramic brackets manufactured by 3D printing: mechanical test of torsion resistance by application of torques | en |
dc.type | Dissertação de mestrado | |
dspace.entity.type | Publication | |
unesp.campus | Universidade Estadual Paulista (UNESP), Instituto de Ciência e Tecnologia, São José dos Campos | pt |
unesp.embargo | Online | pt |
unesp.examinationboard.type | Banca pública | pt |
unesp.graduateProgram | Ciência e Tecnologia Aplicada à Odontologia - ICT | pt |
unesp.knowledgeArea | Inovação tecnológica | pt |
unesp.researchArea | Inovação tecnológica | pt |
Arquivos
Pacote Original
1 - 1 de 1
Carregando...
- Nome:
- silva_pc_me_sjc.pdf
- Tamanho:
- 986.5 KB
- Formato:
- Adobe Portable Document Format
- Descrição:
Licença do Pacote
1 - 1 de 1
Carregando...
- Nome:
- license.txt
- Tamanho:
- 3.01 KB
- Formato:
- Item-specific license agreed upon to submission
- Descrição: