Comparative evaluation of heating instruments for detecting Bacteroides HF183, SARS-CoV-2, and Aichi virus A in untreated wastewater using colorimetric LAMP
Carregando...
Fontes externas
Fontes externas
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
IWA Publishing
Tipo
Artigo
Direito de acesso
Acesso aberto

Fontes externas
Fontes externas
Resumo
This study assessed the accuracy, field suitability, and cost-effectiveness of colorimetric LAMP (c-LAMP) assays using six heating instruments: Electric Kettle, Water Bath, Heating Block, miniPCR, Thermocycler, and Genie® III. Bacteroides HF183 (HF183), SARS-CoV-2, and Aichi virus A (AiV-A) were tested using control materials and wastewater samples, comparing cLAMP with qPCR/RT-qPCR detection. qPCR/RT-qPCR reliably detected HF183 and AiV-A genetic fragments at dilutions as low as 1 × 10<sup>-5</sup> pg/μL, corresponding to 1.24 and 1.49 log<sub>10</sub> GC/reaction. cLAMP/RT-cLAMP detected these fragments at 1 × 10<sup>-4</sup> pg/μL (2.28 and 2.63 log<sub>10</sub> GC/reaction), though Electric Kettle, miniPCR, and Genie® III occasionally produced false negatives. RT-qPCR reliably detected SARS-CoV-2 fragments at 1 × 10<sup>0</sup> GC/μL (1.86 log<sub>10</sub> GC/reaction), with cLAMP/RT-cLAMP performing similarly across most instruments except Thermocycler. No significant differences (p > 0.05) were observed among instruments, though inconsistencies appeared at lower concentrations. The findings offer valuable insights into heating instrument performance for cLAMP/RT-cLAMP assays, guiding their application in wastewater-based pathogen detection.