Publicação:
Thermodynamic response of an ultrasonic densitometer

Nenhuma Miniatura disponível

Data

2005-12-01

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Tipo

Trabalho apresentado em evento

Direito de acesso

Acesso abertoAcesso Aberto

Resumo

An ultrasonic densitometer is experimentally characterized as a function of temperature. The measurement of propagation velocity and density of several liquids is performed in the 15 to 40°C temperature range. Results are compared to tabulated values in the case of distilled water, showing accuracy of 0.07% for the propagation velocity. Density values are compared to those obtained with a pycnometer, showing 0.2% accuracy in density measurement for stabilized temperature and 0.4% under thermal gradient conditions.

Descrição

Idioma

Inglês

Como citar

International Congress on Noise Control Engineering 2005, INTERNOISE 2005, v. 5, p. 4137-4144.