Publicação: Thermodynamic response of an ultrasonic densitometer
Nenhuma Miniatura disponível
Data
2005-12-01
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Tipo
Trabalho apresentado em evento
Direito de acesso
Acesso aberto

Resumo
An ultrasonic densitometer is experimentally characterized as a function of temperature. The measurement of propagation velocity and density of several liquids is performed in the 15 to 40°C temperature range. Results are compared to tabulated values in the case of distilled water, showing accuracy of 0.07% for the propagation velocity. Density values are compared to those obtained with a pycnometer, showing 0.2% accuracy in density measurement for stabilized temperature and 0.4% under thermal gradient conditions.
Descrição
Idioma
Inglês
Como citar
International Congress on Noise Control Engineering 2005, INTERNOISE 2005, v. 5, p. 4137-4144.