BIOSENSORS FOR GLUCOSE NEEDLE-SHAPED FOR IN-VIVO MONITORING
dc.contributor.author | Mascone, D. | |
dc.contributor.author | Yamanaka, H. | |
dc.contributor.author | Mascini, M. | |
dc.contributor.institution | Universidade Estadual Paulista (Unesp) | |
dc.contributor.institution | DIP SANITA PUBBL EPIDEMIOL & CHIM ANALIT AMBIENTALE | |
dc.date.accessioned | 2014-05-20T15:26:26Z | |
dc.date.available | 2014-05-20T15:26:26Z | |
dc.date.issued | 1993-12-01 | |
dc.description.abstract | Different procedures for obtaining a needle biosensor for the determination of glucose to be inserted subcutaneously in vivo, have been compared. Platinum wires with a diameter of 75 mum, teflon-coated were inserted in hypodermic needles and fixed with a two-component epoxy resin. Using a dip-coating procedure, several layers were deposited on electrodes. The first coating was cellulose acetate, the second was immobilized glucose oxidase (GOD) mixed with bovine serum albumin (BSA) and glutaraldheyde, the third coating was a polyurethane coating obtained with commercially available products. A large number of electrodes have been tried and statistically evaluated but they seem to be affected by poor reproducibility evidenced by a large spreading in successive calibration curves. Then, the polyurethane coating has been replaced by a thin polycarbonate membrane salinized and fixed on the tip of the needle. Reproducible results were achieved and first results of in vivo measurements on rabbits are reported. | en |
dc.description.affiliation | UNESP,INST QUIM,BR-14800900 ARARAQUARA,BRAZIL | |
dc.description.affiliation | DIP SANITA PUBBL EPIDEMIOL & CHIM ANALIT AMBIENTALE,SEZ CHIM ANALIT,I-50121 FLORENCE,ITALY | |
dc.description.affiliationUnesp | UNESP,INST QUIM,BR-14800900 ARARAQUARA,BRAZIL | |
dc.format.extent | 1339-1343 | |
dc.identifier | https://getinfo.de/app/Biosensors-for-Glucose-Needle-Shaped-for-in-vivo/id/BLSE%3AEN016629546 | |
dc.identifier.citation | Russian Journal of Electrochemistry. Birmingham: Interperiodica, v. 29, n. 12, p. 1339-1343, 1993. | |
dc.identifier.issn | 1023-1935 | |
dc.identifier.uri | http://hdl.handle.net/11449/36611 | |
dc.identifier.wos | WOS:A1993NY81100017 | |
dc.language.iso | eng | |
dc.publisher | Interperiodica | |
dc.relation.ispartof | Russian Journal of Electrochemistry | |
dc.relation.ispartofjcr | 0.880 | |
dc.relation.ispartofsjr | 0,290 | |
dc.rights.accessRights | Acesso restrito | pt |
dc.source | Web of Science | |
dc.title | BIOSENSORS FOR GLUCOSE NEEDLE-SHAPED FOR IN-VIVO MONITORING | en |
dc.type | Artigo | pt |
dcterms.license | http://www.maik.rssi.ru/cgi-perl/contents.pl?catalog=4&page=1&lang=eng | |
dcterms.rightsHolder | Interperiodica | |
dspace.entity.type | Publication | |
relation.isOrgUnitOfPublication | bc74a1ce-4c4c-4dad-8378-83962d76c4fd | |
relation.isOrgUnitOfPublication.latestForDiscovery | bc74a1ce-4c4c-4dad-8378-83962d76c4fd | |
unesp.campus | Universidade Estadual Paulista (UNESP), Instituto de Química, Araraquara | pt |
Arquivos
Licença do pacote
1 - 1 de 1
Carregando...
- Nome:
- license.txt
- Tamanho:
- 1.71 KB
- Formato:
- Item-specific license agreed upon to submission
- Descrição: