Determination of sulfate in ethanol fuel using an electrode chemically modified with polypyrrole by flow injection analysis
Loading...
Date
Advisor
Coadvisor
Graduate program
Undergraduate course
Journal Title
Journal ISSN
Volume Title
Publisher
Diva Enterprises
Type
Article
Access right
Acesso restrito
Abstract
In this work, an electrode chemically modified with polypyrrole (PCME) was employed for determination of sulfate in ethanol fuel using a FIA system. The PCME was prepared by polymerization of pyrrole at a glassy carbon electrode by means of cyclic voltammetry technique. An analytical curve from 1.0 x 10−5 to 8.0 x 10−5 mol L−1 was obtained in flow injection system based on the PCME. An amperometric sensibility of 2.3 x 10−3 A mol−1 L and a detection limit of 2.5 x 10−6 mol L−1 were achieved. The proposed method was employed for determination of sulfate ions in commercial samples of ethanol fuel. The results were in good agreement with those obtained by the ionic chromatographic method.
Description
Keywords
Language
English
Citation
Journal of Biofuels, v. 1, n. 2, p. 220-227, 2010.