Logotipo do repositório
 

Publicação:
Evaluation and application of bismuth as an internal standard for the determination of lead in wines by simultaneous electrothermal atomic absorption spectrometry

dc.contributor.authorFernandes, K. G.
dc.contributor.authorde Moraes, M.
dc.contributor.authorNeto, JAG
dc.contributor.authorNobrega, J. A.
dc.contributor.authorOliveira, P. V.
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionUniversidade Federal de São Carlos (UFSCar)
dc.contributor.institutionUniversidade de São Paulo (USP)
dc.date.accessioned2014-05-20T15:19:57Z
dc.date.available2014-05-20T15:19:57Z
dc.date.issued2002-01-01
dc.description.abstractA method for the direct determination of Pb in wines by simultaneous multi-element atomic absorption spectrometry (SIMAAS) using a transversely heated graphite atomizer, Zeeman-effect background corrector and internal standardization is proposed. Bismuth was used as an internal standard and Pd(NO3)(2) plus Mg(NO3)(2) as chemical modifier,to stabilize both the analyte and the internal standard. The implementation of two pyrolysis steps avoided any build-up of carbonaceous residues on the graphite platform. All diluted samples (1 + 1 v/v) in 0.2% v/v HNO3 and reference solutions (5.0-50 mug l(-1) Ph in 0.2% v/v HNO3) were spiked with 25 mug l(-1) Bi. For a 20 mul aliquot dispensed into the graphite tube, a good correlation (r = 0.9997) was obtained between the ratio of the analyte signal to the internal standard signal and the analyte concentration of the reference solutions. The electrothermal behaviour of Pb and Bi in red, white and rose wines were compared. In addition, absorbance variations due to changes in experimental conditions, such as atomizer temperature, integration time, injected sample volume, radiation beam intensity, graphite tube surface, dilution and sample composition, were minimized by using Bi as internal standard. Relative standard deviations of measurements based on integrated absorbance varied from 0.1 to 3.4% and from 0.5 to 7.3% (n = 12) with and without internal standard correction, respectively. Good recoveries (91-104%) for Pb spikes were obtained. The characteristic mass was 45 pg Pb and the limit of detection based on integrated absorbance was 0.9 mug l(-1) Pb. Internal standardization increased the lifetime of the tube by 25%. Direct determinations of Pb in wines with and without internal standardization approaches were in agreement at the 95% confidence level. The repeatability and the tube lifetime were improved when using Bi as internal standard. The improvement in accuracy using an internal standard was only observed when the analytical results were affected by errors.en
dc.description.affiliationUniv Estadual Paulista, Dept Quim Analit, BR-14801970 Araraquara, SP, Brazil
dc.description.affiliationUniv Fed Sao Carlos, Dept Quim, BR-13560970 Sao Carlos, SP, Brazil
dc.description.affiliationUniv São Paulo, Inst Quim, BR-05513970 São Paulo, Brazil
dc.description.affiliationUnespUniv Estadual Paulista, Dept Quim Analit, BR-14801970 Araraquara, SP, Brazil
dc.format.extent157-162
dc.identifierhttp://dx.doi.org/10.1039/b105309b
dc.identifier.citationAnalyst. Cambridge: Royal Soc Chemistry, v. 127, n. 1, p. 157-162, 2002.
dc.identifier.doi10.1039/b105309b
dc.identifier.issn0003-2654
dc.identifier.urihttp://hdl.handle.net/11449/31324
dc.identifier.wosWOS:000173475200032
dc.language.isoeng
dc.publisherRoyal Soc Chemistry
dc.relation.ispartofAnalyst
dc.relation.ispartofjcr3.864
dc.relation.ispartofsjr1,249
dc.rights.accessRightsAcesso restrito
dc.sourceWeb of Science
dc.titleEvaluation and application of bismuth as an internal standard for the determination of lead in wines by simultaneous electrothermal atomic absorption spectrometryen
dc.typeArtigo
dcterms.licensehttp://www.rsc.org/AboutUs/Copyright/LicencetoPublishforjournals.asp
dcterms.rightsHolderRoyal Soc Chemistry
dspace.entity.typePublication
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Química, Araraquarapt
unesp.departmentQuímica Analítica - IQARpt

Arquivos

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: