Protocols for in situ measurement of oxygen isotopes in goethite by ion microprobe

dc.contributor.authorMonteiro, H. S. [UNESP]
dc.contributor.authorVasconcelos, P. M.
dc.contributor.authorFarley, K. A.
dc.contributor.authorAvila, J. N.
dc.contributor.authorMiller, H. B. D.
dc.contributor.authorHolden, P.
dc.contributor.authorIreland, T. R.
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionUniv Queensland
dc.contributor.institutionAustralian Natl Univ
dc.contributor.institutionCALTECH
dc.date.accessioned2020-12-10T19:49:26Z
dc.date.available2020-12-10T19:49:26Z
dc.date.issued2020-02-05
dc.description.abstractWe present protocols for high-spatial resolution measurement of oxygen isotope ratios of goethite (alpha-FeOOH) with the Sensitive High Mass Resolution Ion Microprobe - Stable Isotopes (SHRIMP-SI) and propose a natural sample as a potential goethite reference material (RM) for ion microprobe analysis. We assess the effects of goethite chemical composition, crystallographic orientation, and texture on the accuracy and repeatability of SHRIMP-SI delta O-18 (delta O-18(SIMS)) results. Synthetic goethites evaluated as potential delta O-18(SIMS) RM are powdery, porous, and finely crystalline; they do not yield repeatable results. A dense colloform goethite from the Capao topaz mine, Minas Gerais, Brazil, fulfills major prerequisites: it is stoichiometrically relatively pure, yields repeatable oxygen isotope results, and occurs in abundance to produce a RM for long-term use. We use an average laser fluorination delta O-18(LF-VSMOW) value of 17.3 0.3%o (1SD) obtained for five aliquots of this RM to normalize all delta O-18(SIMS) measurements. Multiple delta O-18(SIMS) analyses of a large fragment of the Capao L4 (CL4) RM analyzed in three different runs yield an overall repeatability of -17.3 +/- 0.5 parts per thousand (25D, n = 294) for all three runs combined. Natural variability and crystal orientation effects are the main reasons for the excess spread of the 8180sims results compared to the spot internal precision (ca. 0.2 parts per thousand). All delta O-18(SIMS) analyses (n = 1027) in various aliquots of CL4, randomly oriented and analyzed in 26 sessions during eight distinct runs, yield an overall repeatability of 17.3 +/- 0.7%0 (2SD), confirming that CL4 is a suitable SIMS RM. After ascertaining its suitability as a RM, we used CL4 to standardize analyses of other natural goethite samples with the SHRIMP-SI and compared delta O-18(SIMS) and laser fluorination results to test the relationship between natural properties (e.g., porosity, minor elements substituting for Fe), preparation procedures (e.g., polish and relief), instrument conditions, and the overall precision and accuracy of the SIMS analyses. Samples containing minor elements substituting for Fe (e.g., Al, Mn, Cu, Zn, etc.) or as contaminants (e.g., Si, P) require significant matrix corrections. Because we could not find homogenous natural goethite samples showing a large range in metal concentrations, we extrapolate our calibration curves beyond the composition of our calibration goethite samples. delta O-18(SIMS) results corrected for instrument mass fractionation (using CL4) and compositionally dependent matrix effects (using several calibration goethites of known elemental composition) are less precise but statistically indistinguishable from their laser fluorination results. However, porous samples are unsuitable for SHRIMP-SI delta O-18 analysis. Dense colloform samples yield repeatable results for individual growth bands, showing that the high spatial resolution, moderate precision, and speed of analysis of the SHRIMP-SI can resolve variations in oxygen isotope composition acquired during sample growth. (U-Th)/He geochronology of equivalent aliquots from the same goethite samples reveal that the combination of the two methods permits the extraction of temporal variation in the isotopic compositions of meteoric solutions in the geological past.en
dc.description.affiliationUniv Estadual Paulista, Dept Planejamento Terr & Geoprocessamento, BR-13506900 Rio Claro, SP, Brazil
dc.description.affiliationUniv Queensland, Sch Earth & Environm Sci, Brisbane, Qld 4072, Australia
dc.description.affiliationAustralian Natl Univ, Res Sch Earth Sci, Canberra, ACT 2601, Australia
dc.description.affiliationCALTECH, Div Geol & Planetary Sci, Pasadena, CA 91125 USA
dc.description.affiliationUnespUniv Estadual Paulista, Dept Planejamento Terr & Geoprocessamento, BR-13506900 Rio Claro, SP, Brazil
dc.description.sponsorshipAustralian Research Council (ARC)
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
dc.description.sponsorshipIdAustralian Research Council (ARC): DP160104988
dc.description.sponsorshipIdAustralian Research Council (ARC): LP1401008005
dc.description.sponsorshipIdAustralian Research Council (ARC): LE0560868
dc.format.extent24
dc.identifierhttp://dx.doi.org/10.1016/j.chemgeo.2019.119436
dc.identifier.citationChemical Geology. Amsterdam: Elsevier, v. 533, 24 p., 2020.
dc.identifier.doi10.1016/j.chemgeo.2019.119436
dc.identifier.issn0009-2541
dc.identifier.urihttp://hdl.handle.net/11449/196579
dc.identifier.wosWOS:000513871400025
dc.language.isoeng
dc.publisherElsevier B.V.
dc.relation.ispartofChemical Geology
dc.sourceWeb of Science
dc.subjectGoethite
dc.subjectSHRIMP-SI
dc.subject(U Th)/He
dc.subjectOxygen isotopes
dc.subjectIon microprobe
dc.subjectMatrix effect
dc.titleProtocols for in situ measurement of oxygen isotopes in goethite by ion microprobeen
dc.typeArtigo
dcterms.licensehttp://www.elsevier.com/about/open-access/open-access-policies/article-posting-policy
dcterms.rightsHolderElsevier B.V.

Arquivos