Fracture of Zirconia Abutment with Metallic Insertion on Anterior Single Titanium Implant with Internal Hexagon: Retrieval Analysis of a Failure

dc.contributor.authorBottino, Marco Antonio [UNESP]
dc.contributor.authorVillefort Rocha, Regina Furbino [UNESP]
dc.contributor.authorAnami, Lilian Costa [UNESP]
dc.contributor.authorOzcan, Mutlu
dc.contributor.authorMelo, Renata Marques de [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionUniv Zurich
dc.date.accessioned2018-11-26T17:40:59Z
dc.date.available2018-11-26T17:40:59Z
dc.date.issued2016-09-01
dc.description.abstractThis case report presents a retrieval analysis of a screw-retained one-piece restoration with the veneering ceramic fired directly onto the zirconia abutment that fractured during insertion. A patient who experienced root fracture of a maxillary left central incisor received a titanium implant on the same day as extraction. After delayed implant loading, a two-piece zirconia abutment with metallic insertion was customized. Upon installation, a horizontal fracture of the crown just above the metallic portion was detected. The retrieval analysis of a fractured zirconia abutment showed crack formation and diffusion of glaze material that expanded the crack line during firing.en
dc.description.affiliationSao Paulo State Univ, Inst Sci & Technol, Dept Prosthodont & Dent Mat, Sao Jose Dos Campos, Brazil
dc.description.affiliationUniv Zurich, Dent Mat Unit, Ctr Dent & Oral Med, Clin Fixed & Removable Prosthodont & Dent Mat Sci, Zurich, Switzerland
dc.description.affiliationUnespSao Paulo State Univ, Inst Sci & Technol, Dept Prosthodont & Dent Mat, Sao Jose Dos Campos, Brazil
dc.format.extent164-168
dc.identifierhttp://dx.doi.org/10.1922/EJPRD_01556Bottino05
dc.identifier.citationEuropean Journal Of Prosthodontics And Restorative Dentistry. Lowestoft: Dennis Barber Ltd, v. 24, n. 3, p. 164-168, 2016.
dc.identifier.doi10.1922/EJPRD_01556Bottino05
dc.identifier.issn0965-7452
dc.identifier.lattes9234456003563666
dc.identifier.urihttp://hdl.handle.net/11449/163328
dc.identifier.wosWOS:000411710300007
dc.language.isoeng
dc.publisherDennis Barber Ltd
dc.relation.ispartofEuropean Journal Of Prosthodontics And Restorative Dentistry
dc.relation.ispartofsjr0,265
dc.rights.accessRightsAcesso restrito
dc.sourceWeb of Science
dc.subjectDental Implants
dc.subjectZirconia
dc.subjectProsthesis Failure Analyses
dc.subjectPeriprosthetic Fractures
dc.subjectDental Implant-Abutment Design
dc.subjectDental Porcelain
dc.titleFracture of Zirconia Abutment with Metallic Insertion on Anterior Single Titanium Implant with Internal Hexagon: Retrieval Analysis of a Failureen
dc.typeArtigo
dcterms.rightsHolderDennis Barber Ltd
unesp.advisor.lattes9234456003563666
unesp.author.orcid0000-0003-0077-3161[1]

Arquivos