Evaluation of Lines of Phosphorus and Potassium by High-Resolution Continuum Source Flame Atomic Absorption Spectrometry for Liquid Fertilizer Analysis

dc.contributor.authorFerreira, Roberta Borges [UNESP]
dc.contributor.authorOliveira, Silvana Ruella [UNESP]
dc.contributor.authorFranzini, Vanessa Pezza [UNESP]
dc.contributor.authorVirgilio, Alex [UNESP]
dc.contributor.authorRaposo Junior, Jorge Luiz [UNESP]
dc.contributor.authorGomes Neto, José Anchieta [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2014-05-20T14:19:12Z
dc.date.available2014-05-20T14:19:12Z
dc.date.issued2011-03-01
dc.description.abstractThe usefulness of the side pixel registration approach and molecular absorption was investigated for the determination of phosphorus (P) and potassium (K) in liquid fertilizers by high-resolution continuum source flame atomic absorption spectrometry.The phosphorus oxide (PO) lines at 246.400 nm, 247.620 nm, 247.780 nm, 324.616 nm, and 327.040 nm were evaluated. The influence of side pixel registration to extend the linear working range was also investigated by measuring at wings (404.402 nm, 404.406 nm, 404.410 nm, 404.418 nm, 404.422 nm and 404.426 nm) of the secondary line for K at 404.414 nm. For the lines 327.040 mm and 404.422 nm, analytical curves in the 100-5000 mg L-1 and 50-1000 mg L-1 ranges were obtained with linear correlations better than 0.999 and 0.995, respectively.The proposed method was applied to six liquid fertilizer samples. The found concentrations for K and P were in agreement with those obtained by inductively coupled plasma optical emission spectrometry (ICP-OES), spectrophotometry, and flame atomic emission spectrometry at the 95% confidence level.en
dc.description.affiliationUNESP Univ Estadual Paulista, Dept Analyt Chem, BR-14801970 Araraquara, SP, Brazil
dc.description.affiliationUnespUNESP Univ Estadual Paulista, Dept Analyt Chem, BR-14801970 Araraquara, SP, Brazil
dc.description.sponsorshipFundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
dc.description.sponsorshipIdCNPq: 300600/2009-7
dc.format.extent56-61
dc.identifierhttp://jglobal.jst.go.jp/public/20090422/201202230103386562
dc.identifier.citationAtomic Spectroscopy. Shelton: Perkin-elmer Corp, v. 32, n. 2, p. 56-61, 2011.
dc.identifier.issn0195-5373
dc.identifier.lattes8708735699846578
dc.identifier.urihttp://hdl.handle.net/11449/25787
dc.identifier.wosWOS:000289743800002
dc.language.isoeng
dc.publisherPerkin-elmer Corp
dc.relation.ispartofAtomic Spectroscopy
dc.relation.ispartofjcr1.123
dc.relation.ispartofsjr0,327
dc.rights.accessRightsAcesso restrito
dc.sourceWeb of Science
dc.titleEvaluation of Lines of Phosphorus and Potassium by High-Resolution Continuum Source Flame Atomic Absorption Spectrometry for Liquid Fertilizer Analysisen
dc.typeArtigo
dcterms.rightsHolderPerkin-elmer Corp
unesp.author.lattes8708735699846578
unesp.campusUniversidade Estadual Paulista (Unesp), Instituto de Química, Araraquarapt

Arquivos

Licença do Pacote
Agora exibindo 1 - 2 de 2
Nenhuma Miniatura disponível
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:
Nenhuma Miniatura disponível
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: