Biomechanical evaluation of spring system for implant-supported prosthesis: analysis by photoelasticity and extensometry

dc.contributor.authorGoiato, Marcelo Coelho [UNESP]
dc.contributor.authorde Medeiros, Rodrigo Antonio [UNESP]
dc.contributor.authorda Silva, Emily Vivianne Freitas [UNESP]
dc.contributor.authorSônego, Mariana Vilela [UNESP]
dc.contributor.authordos Santos, Daniela Micheline [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2018-12-11T17:10:52Z
dc.date.available2018-12-11T17:10:52Z
dc.date.issued2017-05-19
dc.description.abstractNew systems are released in the odontological market with the objective of bringing improvements and advancements in the clinical success of implants and implant-supported prostheses. The distribution of tension for the bone/implant system of these new systems is important. The purpose of this work was to evaluate, by photoelastic analysis and strain gauges, the distribution of tension for the bone/implant system, using the Slim system. A photoelastic cast with an external hexagon implant was manufactured and a metal prosthesis was screwed below it, with or without the tested system. For the photoelasticity methodology, a photoelastic cast, with or without the tested system, was positioned in a circular polariscope, and axial loads of 100 N were applied to the centre of the crown. The pattern of tension generated was photographed and analysed qualitatively in an imaging programme. For the extensometry methodology, two extensometers were placed in the mesial and distal region of the implant in the photoelastic cast. The axial loads were applied to the group again, with or without the system (n = 10). The t test of independent samples with a significance level of 5% was used for this analysis. This study demonstrated greater tension values for both the photoelastic and extensometry methods when the tested system was used. Clinical studies must be performed to evaluate the tested system because the results might not be clinically significant to a bone reabsorption.en
dc.description.affiliationDepartment of Dental Materials and Prosthodontics Aracatuba Dental School São Paulo State University (UNESP)
dc.description.affiliationUnespDepartment of Dental Materials and Prosthodontics Aracatuba Dental School São Paulo State University (UNESP)
dc.format.extent309-313
dc.identifierhttp://dx.doi.org/10.1080/03091902.2017.1299230
dc.identifier.citationJournal of Medical Engineering and Technology, v. 41, n. 4, p. 309-313, 2017.
dc.identifier.doi10.1080/03091902.2017.1299230
dc.identifier.file2-s2.0-85016275531.pdf
dc.identifier.issn1464-522X
dc.identifier.issn0309-1902
dc.identifier.lattes9719883814872582
dc.identifier.scopus2-s2.0-85016275531
dc.identifier.urihttp://hdl.handle.net/11449/174386
dc.language.isoeng
dc.relation.ispartofJournal of Medical Engineering and Technology
dc.relation.ispartofsjr0,389
dc.rights.accessRightsAcesso aberto
dc.sourceScopus
dc.subjectbiomechanics
dc.subjectDental implants
dc.subjectphotoelasticity and extensometry
dc.titleBiomechanical evaluation of spring system for implant-supported prosthesis: analysis by photoelasticity and extensometryen
dc.typeArtigo
unesp.author.lattes9719883814872582

Arquivos

Pacote Original
Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
2-s2.0-85016275531.pdf
Tamanho:
690.51 KB
Formato:
Adobe Portable Document Format
Descrição: