Logo do repositório
 

Evaluation and Application of TS-FF-AAS System Using Ceramic Tube as Atomization Cell for Tin Determination in Canned Foods

dc.contributor.authorLobo, Fabiana Aparecida [UNESP]
dc.contributor.authorVillafranca, Ana Cristina [UNESP]
dc.contributor.authorde Oliveira, Adriana Paiva
dc.contributor.authorde Moraes, Mercedes [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionUniv Fed Mato Grosso
dc.date.accessioned2014-05-20T14:19:26Z
dc.date.available2014-05-20T14:19:26Z
dc.date.issued2008-09-01
dc.description.abstractTinplate has been used for preserving food for well over a hundred years. Today it provides a robust form of packaging by allowing minimization of head-space oxygen and sterilization of the foodstuff within the hermetically sealed can. It also provides a long, safe, and ambient shelf life with no or minimal use of preservatives. However, the use of tinplate could result in contamination since some tin could dissolve into the food, particularly when plain uncoated internal surfaces are used. Monitoring this process is, therefore, very important to ensure quality standard and industrial control.This work describes the development of an analytical procedure for on-line tin determination by thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS) using a ceramic tube as an atomization cell in canned foods. Carrier flow rate (air), sample volume, and acid concentration (HCl) were evaluated for the optimization of the TS-FF-AAS system. An analytical linear curve was built [A= 0.0062 + 0.0069 (Sn), (r=0.9993)] in the 3.0-30 mg L-1 Sn range. The limit of detection, relative standard deviation (n=12, 1 mg L-1 Sn), and analytical frequency was: 0.8 mg L-1 Sn, <= 8.5% and 90 h(-1), respectively. Accuracy was checked by performing addition-recovery experiments. Recoveries varied from 95.0 to 104% for nine samples of the canned foods. According to a t-test, all results were compared statistically with values obtained by flame atomic absorption spectrometry (FAAS) using acetylene-nitrous oxide flame and were in agreement at the 95% confidence level.en
dc.description.affiliationUniv Estadual Paulista, UNESP, Dept Quim Analit, BR-14801970 Araraquara, SP, Brazil
dc.description.affiliationUniv Fed Mato Grosso, Dept Quim, BR-78060900 Cuiaba, MT, Brazil
dc.description.affiliationUnespUniv Estadual Paulista, UNESP, Dept Quim Analit, BR-14801970 Araraquara, SP, Brazil
dc.description.sponsorshipFundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
dc.format.extent186-192
dc.identifierhttp://cat.inist.fr/?aModele=afficheN&cpsidt=20839848
dc.identifier.citationAtomic Spectroscopy. Shelton: Perkin-elmer Corp, v. 29, n. 5, p. 186-192, 2008.
dc.identifier.issn0195-5373
dc.identifier.urihttp://hdl.handle.net/11449/25862
dc.identifier.wosWOS:000260705400005
dc.language.isoeng
dc.publisherPerkin-elmer Corp
dc.relation.ispartofAtomic Spectroscopy
dc.relation.ispartofjcr1.123
dc.relation.ispartofsjr0,327
dc.rights.accessRightsAcesso restritopt
dc.sourceWeb of Science
dc.titleEvaluation and Application of TS-FF-AAS System Using Ceramic Tube as Atomization Cell for Tin Determination in Canned Foodsen
dc.typeArtigopt
dcterms.rightsHolderPerkin-elmer Corp
dspace.entity.typePublication
relation.isOrgUnitOfPublicationbc74a1ce-4c4c-4dad-8378-83962d76c4fd
relation.isOrgUnitOfPublication.latestForDiscoverybc74a1ce-4c4c-4dad-8378-83962d76c4fd
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Química, Araraquarapt
unesp.departmentQuímica Analítica - IQARpt

Arquivos

Licença do pacote

Agora exibindo 1 - 2 de 2
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: