Logotipo do repositório
 

Publicação:
Effect of Passive Fit Absence in the Prosthesis/Implant/Retaining Screw System: A Two-Dimensional Finite Element Analysis

dc.contributor.authorGomes, Erica Alves [UNESP]
dc.contributor.authorAssunção, Wirley Goncalves [UNESP]
dc.contributor.authorTabata, Lucas Fernando
dc.contributor.authorRicardo Barao, Valentim Adelino [UNESP]
dc.contributor.authorDelben, Juliana Aparecida [UNESP]
dc.contributor.authorSousa, Edson Antonio Capello [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2013-09-30T18:10:33Z
dc.date.accessioned2014-05-20T13:43:32Z
dc.date.available2013-09-30T18:10:33Z
dc.date.available2014-05-20T13:43:32Z
dc.date.issued2009-11-01
dc.description.abstractThe misfit between prostheses and implants is a clinical reality, but the level that can be accepted without causing mechanical or biologic problem is not well defined. This study investigates the effect of different levels of unilateral angular misfit prostheses in the prosthesis/implant/retaining screw system and in the surrounding bone using finite element analysis. Four models of a two-dimensional finite element were constructed: group I (control), prosthesis that fit the implant; groups 2 to 4, prostheses with unilateral angular misfit of 50, 100, and 200 mu m, respectively. A load of 133 N was applied with a 30-degree angulation and off-axis at 2 mm from the long axis of the implant at the opposite direction of misfit on the models. Taking into account the increase of the angular misfit, the stress maps showed a gradual increase of prosthesis stress and uniform stress in the implant and trabecular bone. Concerning the displacement, an inclination of the system due to loading and misfit was observed. The decrease of the unilateral contact between prosthesis and implant leads to the displacement of the entire system, and distribution and magnitude alterations of the stress also occurred.en
dc.description.affiliationUniv Estadual Paulista, Aracatuba Dent Sch, Dept Dent Mat & Prosthodont, BR-16015050 São Paulo, Brazil
dc.description.affiliationUniv Estadual Paulista, Bauru Engn Sch, Dept Mech Engn, BR-16015050 São Paulo, Brazil
dc.description.affiliationUnespUniv Estadual Paulista, Aracatuba Dent Sch, Dept Dent Mat & Prosthodont, BR-16015050 São Paulo, Brazil
dc.description.affiliationUnespUniv Estadual Paulista, Bauru Engn Sch, Dept Mech Engn, BR-16015050 São Paulo, Brazil
dc.format.extent2000-2005
dc.identifierhttp://dx.doi.org/10.1097/SCS.0b013e3181bd2df8
dc.identifier.citationJournal of Craniofacial Surgery. Philadelphia: Lippincott Williams & Wilkins, v. 20, n. 6, p. 2000-2005, 2009.
dc.identifier.doi10.1097/SCS.0b013e3181bd2df8
dc.identifier.issn1049-2275
dc.identifier.lattes4438747643373395
dc.identifier.urihttp://hdl.handle.net/11449/15200
dc.identifier.wosWOS:000272313600012
dc.language.isoeng
dc.publisherLippincott Williams & Wilkins
dc.relation.ispartofJournal of Craniofacial Surgery
dc.relation.ispartofjcr0.772
dc.relation.ispartofsjr0,448
dc.rights.accessRightsAcesso restritopt
dc.sourceWeb of Science
dc.subjectProsthesisen
dc.subjectmisfiten
dc.subjectbiomechanicsen
dc.subjectdental implantsen
dc.subjectfinite element analysisen
dc.subjectstress distributionen
dc.titleEffect of Passive Fit Absence in the Prosthesis/Implant/Retaining Screw System: A Two-Dimensional Finite Element Analysisen
dc.typeArtigopt
dcterms.licensehttp://edmgr.ovid.com/spine/accounts/copyrightTransfer.pdf
dcterms.rightsHolderLippincott Williams & Wilkins
dspace.entity.typePublication
relation.isOrgUnitOfPublication8b3335a4-1163-438a-a0e2-921a46e0380d
relation.isOrgUnitOfPublication.latestForDiscovery8b3335a4-1163-438a-a0e2-921a46e0380d
unesp.author.lattes8934648262859482
unesp.author.orcid0000-0002-8903-0737[2]
unesp.author.orcid0000-0002-6510-3727[5]
unesp.author.orcid0000-0002-6391-9917[4]
unesp.campusUniversidade Estadual Paulista (UNESP), Faculdade de Engenharia, Baurupt
unesp.campusUniversidade Estadual Paulista (UNESP), Faculdade de Odontologia, Araçatubapt
unesp.departmentEngenharia Mecânica - FEBpt

Arquivos

Licença do Pacote

Agora exibindo 1 - 2 de 2
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: