Publicação:
Validation of a Simplified Implant-Retained Cantilever Fixed Prosthesis

Nenhuma Miniatura disponível

Data

2018-02-01

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Tipo

Artigo

Direito de acesso

Acesso restrito

Resumo

PURPOSE: To evaluate the stress and strain generated in a fixed four-element prosthesis under the application of axial and nonaxial loads using a simplified implant-supported fixed prosthesis model. MATERIALS AND METHODS: A 3-dimensional model was constructed containing 3 implants with a conventional anatomical prosthesis (G1). The second model with the same implant system received the simplified prosthesis (G2). A load of 300 N was applied at an axial point and a nonaxial point through finite element analysis software. RESULTS: The G2 group showed different values of stress concentration in the prosthesis, fixation screw, retention screw, and abutments when compared with G1. Within a limit of 10% degrees of acceptability, the stress on the implants and the bone strain were enclosed for both models of prostheses. CONCLUSION: The simplified fixed prosthesis evaluated presents biomechanical behavior similar to an anatomical prosthesis in the implants and in the surrounding bone structure.

Descrição

Palavras-chave

Idioma

Inglês

Como citar

Implant dentistry, v. 27, n. 1, p. 49-55, 2018.