Publicação: Validation of a Simplified Implant-Retained Cantilever Fixed Prosthesis
Nenhuma Miniatura disponível
Data
2018-02-01
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Tipo
Artigo
Direito de acesso
Acesso restrito
Resumo
PURPOSE: To evaluate the stress and strain generated in a fixed four-element prosthesis under the application of axial and nonaxial loads using a simplified implant-supported fixed prosthesis model. MATERIALS AND METHODS: A 3-dimensional model was constructed containing 3 implants with a conventional anatomical prosthesis (G1). The second model with the same implant system received the simplified prosthesis (G2). A load of 300 N was applied at an axial point and a nonaxial point through finite element analysis software. RESULTS: The G2 group showed different values of stress concentration in the prosthesis, fixation screw, retention screw, and abutments when compared with G1. Within a limit of 10% degrees of acceptability, the stress on the implants and the bone strain were enclosed for both models of prostheses. CONCLUSION: The simplified fixed prosthesis evaluated presents biomechanical behavior similar to an anatomical prosthesis in the implants and in the surrounding bone structure.
Descrição
Palavras-chave
Idioma
Inglês
Como citar
Implant dentistry, v. 27, n. 1, p. 49-55, 2018.