Sensor and biosensor based on Prussian Blue modified gold and platinum screen printed electrodes
Carregando...
Arquivos
Fontes externas
Fontes externas
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Tipo
Artigo
Direito de acesso
Arquivos
Fontes externas
Fontes externas
Resumo
Gold (Au) and platinum (Pt) screen-printed electrodes were modified with Prussian Blue (PB) for the development of amperometric sensors selective for hydrogen peroxide detection. The sensors exhibited sensitivities towards H2O2 equal to 2 A M-1 cm-2 for Au and 1 A M-1 cm-2 for Pt electrodes. The sensors were also employed as the basis for construction of glucose biosensors through further modification with crystallised glucose oxidase immobilised in a Nafion membrane. In order to improve the operational stability of the modified electrodes a buffer solution containing tetrabutylammonium toluene-4-sulfonate was used. The long-term performance of the sensors and biosensors were evaluated by continuous monitoring of hydrogen peroxide and glucose solutions (50 μM and 1 mM, respectively) in the flow-injection mode for 10 h. © 2002 Elsevier Science B.V. All rights reserved.
Descrição
Palavras-chave
Amperometric sensor, Biosensor, Prussian Blue, Screen-printed electrodes
Idioma
Inglês
Citação
Biosensors and Bioelectronics, v. 18, n. 2-3, p. 193-200, 2002.