Logotipo do repositório
 

Publicação:
Determination of copper at wide range concentrations using instrumental features of high-resolution continuum source flame atomic absorption spectrometry

dc.contributor.authorLima, Renata Toledo [UNESP]
dc.contributor.authorRaposo Jr., Jorge Luiz [UNESP]
dc.contributor.authorVirgílio, Alex [UNESP]
dc.contributor.authorGomes Neto, José Anchieta [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2014-05-27T11:25:21Z
dc.date.available2014-05-27T11:25:21Z
dc.date.issued2010-12-01
dc.description.abstractThis work describes a method to determine Cu at wide range concentrations in a single run without need of further dilutions employing high-resolution continuum source flame atomic absorption spectrometry. Different atomic lines for Cu at 324. 754 nm, 327. 396 nm, 222. 570 nm, 249. 215 nm and 224. 426 nm were evaluated and main figures of merit established. Absorbance measurements at 324. 754 nm, 249. 215 nm and 224. 426 nm allows the determination of Cu in the 0. 07-5. 0 mg L -1, 5. 0-100 mg L -1 and 100-800 mg L -1 concentration intervals respectively with linear correlation coefficients better than 0. 998. Limits of detection were 21 μg L -1, 310 μg L -1 and 1400 μg L -1 for 324. 754 nm, 249. 215 nm and 224. 426 nm, respectively and relative standard deviations (n = 12) were ≤ 2. 7%. The proposed method was applied to water samples spiked with Cu and the results were in agreement at a 95% of confidence level (paired t-test) with those obtained by line-source flame atomic absorption spectrometry.en
dc.description.affiliationUnesp - Univ Estadual Paulista Institute of Chemistry, P. O. Box 355, 14801-970 Araraquara - SP
dc.description.affiliationUnespUnesp - Univ Estadual Paulista Institute of Chemistry, P. O. Box 355, 14801-970 Araraquara - SP
dc.format.extent87-92
dc.identifierhttp://dx.doi.org/10.1590/S0100-46702010000400011
dc.identifier.citationEcletica Quimica, v. 35, n. 4, p. 87-92, 2010.
dc.identifier.doi10.1590/S0100-46702010000400011
dc.identifier.file2-s2.0-80052582599.pdf
dc.identifier.issn0100-4670
dc.identifier.lattes8708735699846578
dc.identifier.scieloS0100-46702010000400011
dc.identifier.scopus2-s2.0-80052582599
dc.identifier.urihttp://hdl.handle.net/11449/72074
dc.language.isoeng
dc.relation.ispartofEclética Química
dc.rights.accessRightsAcesso aberto
dc.sourceScopus
dc.subjectCopper
dc.subjectHR-CS FAAS
dc.subjectWide range determination
dc.titleDetermination of copper at wide range concentrations using instrumental features of high-resolution continuum source flame atomic absorption spectrometryen
dc.typeArtigo
dcterms.licensehttp://www.scielo.br/revistas/eq/paboutj.htm
dspace.entity.typePublication
unesp.author.lattes8708735699846578
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Química, Araraquarapt
unesp.departmentQuímica Analítica - IQARpt

Arquivos

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
2-s2.0-80052582599.pdf
Tamanho:
139.58 KB
Formato:
Adobe Portable Document Format