Accuracy of mechanical torque devices for implants used in Brazilian dental offices

Carregando...
Imagem de Miniatura

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Tipo

Artigo

Direito de acesso

Resumo

The purpose of this study was to determine the accuracy of mechanical torque devices in delivering target torque values in dental offices in Salvador, Brazil. A team of researchers visited 16 dental offices, and the clinicians applied torque values (20 and 32 Ncm) to electronic torque controllers. Five repetitions were completed at each torque value and data were collected. When 20 Ncm of torque was used, 62.5% of measured values were accurate (within 10% of the target value). For 32 Ncm, however, only 37.5% of these values were achieved. Several of the tested mechanical torque devices were inaccurate. © 2011 by Quintessence Publishing Co Inc.

Descrição

Palavras-chave

Idioma

Inglês

Como citar

International Journal of Prosthodontics, v. 24, n. 1, p. 38-39, 2011.