Publicação: Digital controller design considering hardware constraints: application in a paraplegic patient
dc.contributor.author | Sanches, Marcelo Augusto Assunção [UNESP] | |
dc.contributor.author | Gaino, Ruberlei | |
dc.contributor.author | Kozan, Renan Fernandes [UNESP] | |
dc.contributor.author | Teixeira, Marcelo Carvalho Minhoto [UNESP] | |
dc.contributor.author | Carvalho, Aparecido Augusto De [UNESP] | |
dc.contributor.author | Covacic, Márcio Roberto | |
dc.contributor.author | Alves, Carlos Antonio [UNESP] | |
dc.contributor.author | Urban, Mateus Fernandes Réu [UNESP] | |
dc.contributor.author | Junqueira, Marcos Vinícius Nascimento [UNESP] | |
dc.contributor.author | Cardim, Rodrigo [UNESP] | |
dc.contributor.author | Assunção, Edvaldo [UNESP] | |
dc.contributor.author | Gentilho Junior, Edno | |
dc.contributor.institution | Universidade Estadual Paulista (Unesp) | |
dc.contributor.institution | Londrina State University Center of Technology and Urbanism Department of Electrical Engineering | |
dc.date.accessioned | 2015-02-02T12:39:27Z | |
dc.date.available | 2015-02-02T12:39:27Z | |
dc.date.issued | 2014-09-01 | |
dc.description.abstract | INTRODUCTION: A methodology was developed for implementing closed-loop control algorithms and for evaluating the behavior of a system, considering certain component restrictions used in laboratory implementation. METHODS: Mathematical functions representing a model of the biological system were used for knee extension/flexion movements. A Proportional Integral Derivative (PID) controller and another one using the root locus method were designed to control a patient’s leg position by applying functional electrical stimulation (FES). The controllers were simulated in Matlab and ISIS Proteus. After the simulations were performed, the codes were embedded in a microcontroller, and tests were conducted on a paraplegic volunteer. To the best of the authors’ knowledge, this is the first time that ISIS Proteus software resources have been used prior to implementing a closed-loop system designed to control the leg position of patients. RESULTS:This method obviates the application of initial controller tests directly to patients. The response obtained in the experiment with a paraplegic patient complied with the specifications set in terms of the steady-state error, the settling time, and the percentage overshoot. The proposed procedure was successfully applied for the implementation of a controller used to control the leg position of a paraplegic person by electrical muscle stimulation. CONCLUSION:The methodology presented in this manuscript can contribute to the implementation of analog and digital controllers because hardware limitations are typically not taken into account in the design of controllers. | en |
dc.description.affiliation | Universidade Estadual Paulista Department of Electrical Engineering | |
dc.description.affiliation | Londrina State University Center of Technology and Urbanism Department of Electrical Engineering | |
dc.description.affiliationUnesp | Universidade Estadual Paulista Department of Electrical Engineering | |
dc.description.sponsorship | Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP) | |
dc.description.sponsorship | Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) | |
dc.description.sponsorship | Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) | |
dc.format.extent | 232-241 | |
dc.identifier | http://dx.doi.org/10.1590/1517-3151.0196 | |
dc.identifier.citation | Revista Brasileira de Engenharia Biomédica. SBEB - Sociedade Brasileira de Engenharia Biomédica, v. 30, n. 3, p. 232-241, 2014. | |
dc.identifier.doi | 10.1590/1517-3151.0196 | |
dc.identifier.file | S1517-31512014000300005.pdf | |
dc.identifier.issn | 1517-3151 | |
dc.identifier.lattes | 0250066159980825 | |
dc.identifier.lattes | 8755160580142626 | |
dc.identifier.lattes | 8630973561889272 | |
dc.identifier.lattes | 8879964582778840 | |
dc.identifier.lattes | 5062087380571462 | |
dc.identifier.orcid | 0000-0002-1072-3814 | |
dc.identifier.scielo | S1517-31512014000300005 | |
dc.identifier.uri | http://hdl.handle.net/11449/114345 | |
dc.language.iso | eng | |
dc.publisher | SBEB - Sociedade Brasileira de Engenharia Biomédica | |
dc.relation.ispartof | Revista Brasileira de Engenharia Biomédica | |
dc.relation.ispartofsjr | 0,179 | |
dc.rights.accessRights | Acesso aberto | |
dc.source | SciELO | |
dc.subject | Digital controller | en |
dc.subject | Microcontroller | en |
dc.subject | ISIS Proteus | en |
dc.subject | PID controller | en |
dc.subject | Functional Electrical Stimulation (FES) | en |
dc.subject | Paraplegic | en |
dc.title | Digital controller design considering hardware constraints: application in a paraplegic patient | en |
dc.type | Artigo | |
dspace.entity.type | Publication | |
unesp.author.lattes | 0250066159980825[5] | |
unesp.author.lattes | 8755160580142626 | |
unesp.author.lattes | 8630973561889272 | |
unesp.author.lattes | 8879964582778840 | |
unesp.author.lattes | 7296886099415451[7] | |
unesp.author.lattes | 5062087380571462[10] | |
unesp.author.orcid | 0000-0001-8204-3482[5] | |
unesp.author.orcid | 0000-0002-1072-3814[10] | |
unesp.campus | Universidade Estadual Paulista (UNESP), Faculdade de Engenharia, Ilha Solteira | pt |
unesp.department | Engenharia Elétrica - FEIS | pt |
Arquivos
Pacote Original
1 - 1 de 1
Carregando...
- Nome:
- S1517-31512014000300005.pdf
- Tamanho:
- 861.13 KB
- Formato:
- Adobe Portable Document Format