Publicação: Validation of an HPLC method for quantitative determination of Benzocaine in PHBV-microparticles and PLA-nanoparticles
Nenhuma Miniatura disponível
Data
2009-08-20
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Tipo
Artigo
Direito de acesso
Resumo
Benzocaine (BZC) is an ester-type local anesthetic used mainly in topical, dermal and mucosal formulations. The present work consists of the development and validation of analytical methodology for evaluation of benzocaine content in the micro and nanoparticles produced by biodegradable polymers (polyhydroxybutyrate-co-hydroxyvalerate and poly-(L-lactide) by HPLC. The validation was done using the reversed-phase C18 column, using a mobile phase consisting of acetonitrile/water 50:50 (v/v), flow rate of 1.5 mL/min and UV-vis detector at 285 nm. The results here obtained showed that the analytical methodology is accurate, reproducible, robust and linear over the molar range concentration of 10-100 μM of benzocaine. The limit of quantification and detection was 13.06 μM and 3.92 μM, respectively. The encapsulation efficiency of benzocaine in PHBV microspheres and PLA nanocapsule were 40% and 70%, respectively. The method developed was applied in the analysis of benzocaine in micro and nanoparticle systems and showed to be efficient, yielding good results. Here, this method was used to evaluate the encapsulation efficiency of benzocaine and will be used in next studies with different micro/ nanoparticle formulations.
Descrição
Palavras-chave
Idioma
Inglês
Como citar
Latin American Journal of Pharmacy, v. 28, n. 3, p. 393-399, 2009.