A device for rapid calorimetric measurements on small biological tissue samples
Carregando...
Arquivos
Fontes externas
Fontes externas
Data
Orientador
Coorientador
Pós-graduação
Curso de graduação
Título da Revista
ISSN da Revista
Título de Volume
Editor
Tipo
Artigo
Direito de acesso
Arquivos
Fontes externas
Fontes externas
Resumo
A new calorimetric technique is described that allows high-throughput heat production rate measurements on small biological tissue samples. The technique is based on the widely used thermopile chip technology combined with an innovative method for precise transport and positioning of samples of different biological materials at the thermal power detector inside the calorimeter. The new transport and positioning technique is a combination of fluidic and mechanical transport, where the latter is realized by a magneto-motor drive. The transport facility ensures good diffusive oxygen penetration into the sample, which is essential for highly metabolically active materials. The proper functioning of the device is demonstrated by measuring the heat production of metabolically active brown adipose tissue, biopsied tegu lizard muscle, and live Drosophila larvae at different stages and temperatures.
Descrição
Palavras-chave
Biological tissues, Brown adipose tissue, Chip calorimetry, Drosophila larvae, Tegu lizard muscle tissue
Idioma
Inglês
Citação
Journal of Thermal Analysis and Calorimetry, v. 149, n. 15, p. 8085-8096, 2024.