A device for rapid calorimetric measurements on small biological tissue samples

Carregando...
Imagem de Miniatura

Orientador

Coorientador

Pós-graduação

Curso de graduação

Título da Revista

ISSN da Revista

Título de Volume

Editor

Tipo

Artigo

Direito de acesso

Resumo

A new calorimetric technique is described that allows high-throughput heat production rate measurements on small biological tissue samples. The technique is based on the widely used thermopile chip technology combined with an innovative method for precise transport and positioning of samples of different biological materials at the thermal power detector inside the calorimeter. The new transport and positioning technique is a combination of fluidic and mechanical transport, where the latter is realized by a magneto-motor drive. The transport facility ensures good diffusive oxygen penetration into the sample, which is essential for highly metabolically active materials. The proper functioning of the device is demonstrated by measuring the heat production of metabolically active brown adipose tissue, biopsied tegu lizard muscle, and live Drosophila larvae at different stages and temperatures.

Descrição

Palavras-chave

Biological tissues, Brown adipose tissue, Chip calorimetry, Drosophila larvae, Tegu lizard muscle tissue