Potentiometric study using chemically modified silica gel with pyridinium as membrane for ClO4 - ions
Loading...
Files
External sources
External sources
Date
Advisor
Coadvisor
Graduate program
Undergraduate course
Journal Title
Journal ISSN
Volume Title
Publisher
Type
Article
Access right
Acesso restrito
Files
External sources
External sources
Abstract
A potentiometric sensor for the perchlorate anion was developed by mixing chemically modified silicagel with pyridinium perchlorate, with an epoxy polymer and graphite. The electrode showed Nernstian response between 1.0 × 10-2 and 1.0 × 10-3M perchlorate concentrations. The electrode showed high selectivity to this ion at solutions pH between 5.5 and 8.0. The presence of IO4 -, NO3 -,Br-, IO3 -, Cl- and SO4 2- ions in the solutions, had only small interference in the electrode response in the range mentioned. © 1995 Springer-Verlag.
Description
Keywords
modified silica gel, perchlorate analysis, perchlorate selective electrode, pyridinium ion
Language
English
Citation