Publicação: Absorbance detector for high performance liquid chromatography based on a deep-UV light-emitting diode at 235 nm
dc.contributor.author | da Silveira Petruci, João Flavio [UNESP] | |
dc.contributor.author | Liebetanz, Michael G. | |
dc.contributor.author | Cardoso, Arnaldo Alves [UNESP] | |
dc.contributor.author | Hauser, Peter C. | |
dc.contributor.institution | University of Basel | |
dc.contributor.institution | Universidade Estadual Paulista (Unesp) | |
dc.date.accessioned | 2018-12-11T17:33:13Z | |
dc.date.available | 2018-12-11T17:33:13Z | |
dc.date.issued | 2017-08-25 | |
dc.description.abstract | In this communication, we describe a flow-through optical absorption detector for HPLC using for the first time a deep-UV light-emitting diode with an emission band at 235 nm as light source. The detector is also comprised of a UV-sensitive photodiode positioned to enable measurement of radiation through a flow-through cuvette with round aperture of 1 mm diameter and optical path length of 10 mm, and a second one positioned as reference photodiode; a beam splitter and a power supply. The absorbance was measured and related to the analyte concentration by emulating the Lambert-Beer law with a log-ratio amplifier circuitry. This detector showed noise levels of 0.30 mAU, which is comparable with our previous LED-based detectors employing LEDs at 280 and 255 nm. The detector was coupled to a HPLC system and successfully evaluated for the determination of the anti-diabetic drugs pioglitazone and glimepiride in an isocratic separation and the benzodiazepines flurazepam, oxazepam and clobazam in a gradient elution. Good linearities (r > 0.99), a precision better than 0.85% and limits of detection at sub-ppm levels were achieved. | en |
dc.description.affiliation | Department of Chemistry University of Basel, Spitalstrasse 51 | |
dc.description.affiliation | São Paulo State University (UNESP) Department of Analytical Chemistry | |
dc.description.affiliationUnesp | São Paulo State University (UNESP) Department of Analytical Chemistry | |
dc.format.extent | 143-146 | |
dc.identifier | http://dx.doi.org/10.1016/j.chroma.2017.07.029 | |
dc.identifier.citation | Journal of Chromatography A, v. 1512, p. 143-146. | |
dc.identifier.doi | 10.1016/j.chroma.2017.07.029 | |
dc.identifier.file | 2-s2.0-85023647551.pdf | |
dc.identifier.issn | 1873-3778 | |
dc.identifier.issn | 0021-9673 | |
dc.identifier.scopus | 2-s2.0-85023647551 | |
dc.identifier.uri | http://hdl.handle.net/11449/179025 | |
dc.language.iso | eng | |
dc.relation.ispartof | Journal of Chromatography A | |
dc.relation.ispartofsjr | 1,378 | |
dc.rights.accessRights | Acesso aberto | |
dc.source | Scopus | |
dc.subject | Absorbance detection | |
dc.subject | Deep-UV | |
dc.subject | HPLC | |
dc.subject | Light-emitting diode | |
dc.title | Absorbance detector for high performance liquid chromatography based on a deep-UV light-emitting diode at 235 nm | en |
dc.type | Artigo | |
dspace.entity.type | Publication | |
unesp.author.lattes | 9165109840414837[3] | |
unesp.author.orcid | 0000-0003-2046-995X[3] | |
unesp.campus | Universidade Estadual Paulista (UNESP), Instituto de Química, Araraquara | pt |
unesp.department | Química Analítica - IQAR | pt |
Arquivos
Pacote Original
1 - 1 de 1
Carregando...
- Nome:
- 2-s2.0-85023647551.pdf
- Tamanho:
- 690.22 KB
- Formato:
- Adobe Portable Document Format
- Descrição: