Cathodic stripping voltammetric determination of cefaclor in pharmaceutical formulations
Loading...
Files
External sources
External sources
Date
Advisor
Coadvisor
Graduate program
Undergraduate course
Journal Title
Journal ISSN
Volume Title
Publisher
Marcel Dekker Inc
Type
Article
Access right
Acesso restrito
Files
External sources
External sources
Abstract
A sensitive method is described for the determination of cefaclor by cathodic stripping voltammetry at the hanging mercury drop electrode. cefaclor is accumulated at the electrode surface as a mercury salt, which is reduced at -0.67 V. The optimum accumulation potential and accumulation time were +0.15 V and up to 180 s, respectively. Linear calibration graphs were obtained between 3.9 mu g.L-1 to 39 mu g.L-1 and the limit of determination was evaluated to be 1.9 mu g.L-1. The method was applied successfully to the determination of cefaclor in pharmaceutical formulations.
Description
Keywords
cefaclor, pharmaceuticals, stripping voltammetry, hanging mercury drop
Language
English
Citation
Analytical Letters. New York: Marcel Dekker Inc., v. 32, n. 1, p. 97-109, 1999.