Efeito do hélio na resistência ao fluxo de gás durante a ventilação mecânica: simulação in vitro para gatos
| dc.contributor.advisor | Teixeira Neto, Francisco José [UNESP] | |
| dc.contributor.author | Chacon, Weslei Souza [UNESP] | |
| dc.contributor.committeeMember | Teixeira Neto, Francisco José [UNESP] | |
| dc.contributor.committeeMember | Nascimento Junior, Paulo do [UNESP] | |
| dc.contributor.committeeMember | Fantoni, Denise Tabacchi | |
| dc.contributor.institution | Universidade Estadual Paulista (Unesp) | pt |
| dc.date.accessioned | 2026-06-30T14:00:04Z | |
| dc.date.issued | 2025-07-30 | |
| dc.description.abstract | Objetivo: Avaliar os efeitos de diferentes misturas de hélio/oxigênio na resistência inspiratória (Rinsp) e expiratória (Rexp) ao fluxo de gás, utilizando um modelo in vitro de ventilação controlada por volume em gatos. Material e Métodos: Um ventilador mecânico foi ajustado para fornecer um volume corrente expirado de 40 mL (tempo inspiratório: 1,5 segundos; pausa inspiratória: 0,75 segundos; pressão expiratória final: 0 cmH2O). O ventilador foi conectado a um pulmão de teste por meio de tubos endotraqueais com diferentes diâmetros internos (n = 6 para cada diâmetro interno), cujo comprimento foi padronizado para 14 cm. Uma faixa elástica foi colocada ao redor do pulmão de teste para manter a complacência em 8 mL/cmH2O. A Rinsp e a Rexp foram medidas durante reduções progressivas no diâmetro interno do tubo endotraqueal (ETtube(ID)), de 4,5 para 3,5 e 2,5 mm, utilizando uma concentração inspirada de oxigênio de 100% (FIO2(100%)), misturas de nitrogênio/oxigênio (50% e 30% de FIO2), misturas de 50% de hélio/50% de oxigênio (Heliox(50%)) e 70% de hélio/30% de oxigênio (Heliox(70%)). Uma ANOVA bidirecional seguida por testes de Dunnet ou Bonferroni comparou os efeitos da redução do ETtube(ID) e das misturas de gases na resistência ao fluxo de gás, respectivamente (p < 0,05). Resultados: À medida que o ETtube(ID) foi reduzido de 4,5 para 2,5 mm, a Rinsp e a Rexp aumentaram significativamente, independentemente da mistura de gases (p < 0,0001 a p = 0,0007). Para todos os tubos ETtube(ID), as misturas de hélio/oxigênio diminuíram significativamente a Rinsp e a Rexp em comparação com a FIO2(100%) e com as misturas de nitrogênio/oxigênio (p < 0,0001 a p = 0,016). Como a Rinsp aumentou 10 vezes com a redução do ETtube(ID) de 4,5 para 2,5 mm com hélio/oxigênio, as reduções em Rinsp e Rexp foram significativamente maiores com Heliox(70%) do que com Heliox(50%) (p < 0,0001). Conclusões e relevância clínica: As misturas de hélio/oxigênio minimizaram o aumento da resistência ao fluxo de gás induzido pela diminuição progressiva do ETtube(ID). O efeito benéfico no aumento da concentração de hélio de 50 para 70% foi maior quando a resistência ao fluxo de gás foi maior. | pt |
| dc.description.abstract | Objective To evaluate the effects of different helium/oxygen mixtures on the inspiratory (Rinsp) and expiratory (Rexp) resistance to gas flow using an in vitro model of volumecontrolled ventilation in cats. Material and Methods A mechanical ventilator was adjusted to deliver an expired tidal volume of 40 mL (inspiratory time, 1.5 seconds; inspiratory pause, 0.75 seconds; endexpiratory pressure, 0 cmH2O). The ventilator was connected to a test lung though endotracheal tubes with different internal diameters (n = 6 for each internal diameter), whose length was standardized to 14 cm. An elastic band was placed around the test lung to maintain compliance at 8 mL/cmH2O. The Rinsp and Rexp were measured during progressive decreases in endotracheal tube internal diameter (ETtube(ID)), from 4.5 to 3.5 and 2.5 mm, using an inspired oxygen concentration of 100% (FIO2(100%)), nitrogen/oxygen mixtures (50% and 30% FiO2), 50% helium/50% oxygen (Heliox(50%)), 70% helium/30% oxygen (Heliox(70%)) mixtures. A two-way ANOVA followed by a Dunnet´s or a Bonferroni compared the effects of decreasing ETtube(ID) and of gas mixtures on resistance to gas flow, respectively (p < 0.05) Results As the ETtube(ID) was reduced from 4.5 to 2.5 mm, Rinsp and Rexp were significantly increased, regardless of the gas mixture (p < 0.0001 to p = 0.0007). For all ETtube(IDs), helium/oxygen mixtures significantly decreased Rinsp and Rexp in comparison with FIO2(100%) and with nitrogen/oxygen mixtures (p < 0.0001 to p = 0.016). As Rinsp was increased by 10-fold by reducing ETtube(ID) from 4.5 to 2.5 mm with helium/oxygen, decreases in Rinsp and Rexp where significantly Heliox(70%) than with Heliox(50%) (p < 0.0001). Conclusions and clinical relevance Helium/oxygen mixtures minimized the increase in resistance to gas flow induced by progressive decreases in ETtube(ID). The beneficial effect in increasing the helium concentration from 50 to 70% was greater when resistance to gas flow was higher. | en |
| dc.description.sponsorship | Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) | |
| dc.description.sponsorshipId | CAPES: 88887901308/2023-00 | |
| dc.identifier.capes | 33004064076P6 | |
| dc.identifier.citation | CHACON, Weslei Souza. Efeito do hélio na resistência ao fluxo de gás durante a ventilação mecânica: simulação in vitro para gatos. 2026. Dissertação (Mestrado em Anestesiologia) – Faculdade de Medicina, Universidade Estadual Paulista (UNESP), Botucatu, 2025. | |
| dc.identifier.lattes | 4988711956208910 | |
| dc.identifier.orcid | 0009-0001-0923-8694 | |
| dc.identifier.uri | https://hdl.handle.net/11449/326859 | |
| dc.language.iso | por | |
| dc.publisher | Universidade Estadual Paulista (Unesp) | |
| dc.rights.accessRights | Acesso aberto | pt |
| dc.subject | Pressão positiva contínua nas vias aéreas | pt |
| dc.subject | Heliox | pt |
| dc.subject | Gatos | pt |
| dc.subject | Mecânica pulmonar | pt |
| dc.subject | Respiração artificial | pt |
| dc.title | Efeito do hélio na resistência ao fluxo de gás durante a ventilação mecânica: simulação in vitro para gatos | pt |
| dc.title.alternative | Effect of helium on gas flow resistance during mechanical ventilation: an in vitro simulation for cats | en |
| dc.type | Dissertação de mestrado | pt |
| dspace.entity.type | Publication | |
| relation.isGradProgramOfPublication | c78518f9-6cc1-430a-813b-d8879d0131d3 | |
| relation.isGradProgramOfPublication.latestForDiscovery | c78518f9-6cc1-430a-813b-d8879d0131d3 | |
| relation.isOrgUnitOfPublication | a3cdb24b-db92-40d9-b3af-2eacecf9f2ba | |
| relation.isOrgUnitOfPublication.latestForDiscovery | a3cdb24b-db92-40d9-b3af-2eacecf9f2ba | |
| unesp.campus | Universidade Estadual Paulista (UNESP), Faculdade de Medicina, Botucatu | pt |
| unesp.examinationboard.type | Banca pública | pt |
| unesp.graduateProgram | Anestesiologia - FMB | pt |
| unesp.knowledgeArea | Risco e proteção de órgãos e sistemas em anestesia e cirurgia | pt |
| unesp.researchArea | Efeitos de fármacos e técnicas anestésicas em órgãos e sistemas | pt |
Arquivos
Pacote original
1 - 1 de 1
Carregando...
- Nome:
- chacon_ws_me_bot.pdf
- Tamanho:
- 1,68 MB
- Formato:
- Adobe Portable Document Format

