Publicação:
Influence of the Connector and Implant Design on the Implant-Tooth-Connected Prostheses

dc.contributor.authorda Silva, Edmar Ferreira
dc.contributor.authorPellizzer, Eduardo Piza [UNESP]
dc.contributor.authorMazaro, José Vitor Quinelli [UNESP]
dc.contributor.authorGarcia, Idelmo Rangel [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2013-09-30T18:30:06Z
dc.date.accessioned2014-05-20T13:43:11Z
dc.date.available2013-09-30T18:30:06Z
dc.date.available2014-05-20T13:43:11Z
dc.date.issued2010-01-01
dc.description.abstractPurpose:The purpose of this study was to evaluate stress transfer patterns between implant-tooth-connected prostheses comparing rigid and semirigid connectors and internal and external hexagon implants.Materials and Methods:Two models were made of photoelastic resin PL-2, with an internal hexagon implant of 4.00 x 13 mm and another with an external hexagon implant of 4.00 x 13 mm. Three denture designs were fabricated for each implant model, incorporating one type of connection in each one to connect implants and teeth: 1) welded rigid connection; 2) semirigid connection; and 3) rigid connection with occlusal screw. The models were placed in the polariscope, and 100-N axial forces were applied on fixed points on the occlusal surface of the dentures.Results:There was a trend toward less intensity in the stresses on the semirigid connection and solid rigid connection in the model with the external hexagon; among the three types of connections in the model with the internal hexagon implant, the semirigid connection was the most unfavorable one; in the tooth-implant association, it is preferable to use the external hexagon implant.Conclusions:The internal hexagon implant establishes a greater depth of hexagon retention and an increase in the level of denture stability in comparison with the implant with the external hexagon. However, this greater stability of the internal hexagon generated greater stresses in the abutment structures. Therefore, when this association is necessary, it is preferable to use the external hexagon implant.en
dc.description.affiliationSão Paulo State Univ Aracatuba, Dept Dent Mat & Prostheses, Dept Dent Mat & Prosthodont, BR-16015050 São Paulo, Brazil
dc.description.affiliationSão Paulo State Univ Aracatuba, Surg & Integrated Clin Dept, BR-16015050 São Paulo, Brazil
dc.description.affiliationUnespSão Paulo State Univ Aracatuba, Dept Dent Mat & Prostheses, Dept Dent Mat & Prosthodont, BR-16015050 São Paulo, Brazil
dc.description.affiliationUnespSão Paulo State Univ Aracatuba, Surg & Integrated Clin Dept, BR-16015050 São Paulo, Brazil
dc.format.extent254-262
dc.identifierhttp://dx.doi.org/10.1111/j.1708-8208.2009.00161.x
dc.identifier.citationClinical Implant Dentistry and Related Research. Malden: Wiley-blackwell, v. 12, n. 3, p. 254-262, 2010.
dc.identifier.doi10.1111/j.1708-8208.2009.00161.x
dc.identifier.issn1523-0899
dc.identifier.lattes5581364193525500
dc.identifier.urihttp://hdl.handle.net/11449/15036
dc.identifier.wosWOS:000280629900011
dc.language.isoeng
dc.publisherWiley-Blackwell
dc.relation.ispartofClinical Implant Dentistry and Related Research
dc.relation.ispartofjcr3.097
dc.relation.ispartofsjr1,827
dc.rights.accessRightsAcesso restritopt
dc.sourceWeb of Science
dc.subjectdental implanten
dc.subjectimplant designen
dc.subjectphotoelasticityen
dc.subjecttooth-implant connectionen
dc.titleInfluence of the Connector and Implant Design on the Implant-Tooth-Connected Prosthesesen
dc.typeArtigopt
dcterms.licensehttp://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1601-0825/homepage/ForAuthors.html
dcterms.rightsHolderWiley-blackwell
dspace.entity.typePublication
relation.isOrgUnitOfPublication8b3335a4-1163-438a-a0e2-921a46e0380d
relation.isOrgUnitOfPublication.latestForDiscovery8b3335a4-1163-438a-a0e2-921a46e0380d
unesp.author.lattes5581364193525500
unesp.author.orcid0000-0003-0670-5004[2]
unesp.campusUniversidade Estadual Paulista (UNESP), Faculdade de Odontologia, Araçatubapt
unesp.departmentCirurgia e Clínica Integrada - FOApt

Arquivos

Licença do Pacote

Agora exibindo 1 - 2 de 2
Nenhuma Miniatura disponível
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:
Nenhuma Miniatura disponível
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: