Logotipo do repositório
 

Publicação:
Determination of cadmium in river water samples by flame AAS after on-line preconcentration in mini-column packed with 2-aminothiazole-modified silica gel

dc.contributor.authorAlcântara, Ilton L. [UNESP]
dc.contributor.authorRoldan, Paulo S. [UNESP]
dc.contributor.authorCastro, Gustavo R. [UNESP]
dc.contributor.authorMoraes, Fabrício V. [UNESP]
dc.contributor.authorSilva, Fábio A. [UNESP]
dc.contributor.authorPadilha, Cilene C. F. [UNESP]
dc.contributor.authorOliveira, Jorge D. [UNESP]
dc.contributor.authorPadilha, Pedro M. [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionUniversidade Estadual do Maranhão (UEMA)
dc.date.accessioned2014-05-27T11:21:06Z
dc.date.available2014-05-27T11:21:06Z
dc.date.issued2004-07-01
dc.description.abstractA rapid and sensitive method was developed to determine trace levels of Cd2+ ions in an aqueous medium by flame atomic absorption spectrometry, using on-line preconcentration in a mini-column packed with 100 mg of 2-aminothiazol modified silica gel (SiAT). The Cd2+ ions were sorbed at pH 5.0. The preconcentrated Cd2+ ions were directly eluted from the column to the spectrometer's nebulizer-burner system using 100 μL of 2 mol L-1 hydrochloric acid. A retention efficiency of over 95% was achieved. The enrichment factor (calculated as the ratio of slopes of the calibration graphs) obtained with preconcentrations in a mini-column packed with SiAT (A = -1.3 × 10-3 + 1.8 × 10-3 [Cd2+]) and without preconcentrations (A = 4 × 10-5 + 3.5 × 10-3[Cd2+]), was 51 and the detection limit calculated was 0.38 μg L-1. The preconcentration procedure was applied to determine trace levels of Cd in river water samples. The optimum preconcentration conditions are discussed herein.en
dc.description.affiliationInstituto de Química Departamento de Quimica Analitica UNESP, P. O. Box 355, 14800-900, Araraquara, SP
dc.description.affiliationInstituto de Biociência Departamento de Quimica e Bioquimica UNESP, P. O. Box 510, 18618-000, Botucatu, SP
dc.description.affiliationInstituto de Biociências Departamento de Fisica e Biofisica UNESP, P. O. Box 510, 18618-000, Botucatu, SP
dc.description.affiliationUniversidade Estadual do Maranhao CESI, 65901-480, Imperatriz, MA
dc.description.affiliationUnespInstituto de Química Departamento de Quimica Analitica UNESP, P. O. Box 355, 14800-900, Araraquara, SP
dc.description.affiliationUnespInstituto de Biociência Departamento de Quimica e Bioquimica UNESP, P. O. Box 510, 18618-000, Botucatu, SP
dc.description.affiliationUnespInstituto de Biociências Departamento de Fisica e Biofisica UNESP, P. O. Box 510, 18618-000, Botucatu, SP
dc.format.extent1029-1032
dc.identifierhttp://dx.doi.org/10.2116/analsci.20.1029
dc.identifier.citationAnalytical Sciences, v. 20, n. 7, p. 1029-1032, 2004.
dc.identifier.doi10.2116/analsci.20.1029
dc.identifier.file2-s2.0-3542992106.pdf
dc.identifier.issn0910-6340
dc.identifier.lattes6981448637456391
dc.identifier.orcid0000-0003-4179-0574
dc.identifier.scopus2-s2.0-3542992106
dc.identifier.urihttp://hdl.handle.net/11449/67792
dc.identifier.wosWOS:000222697500007
dc.language.isoeng
dc.relation.ispartofAnalytical Sciences
dc.relation.ispartofjcr1.355
dc.relation.ispartofsjr0,424
dc.rights.accessRightsAcesso aberto
dc.sourceScopus
dc.subject2 aminothiazole derivative
dc.subjectcadmium
dc.subjecthydrochloric acid
dc.subjectriver water
dc.subjectsilica gel
dc.subjectadsorption
dc.subjectaqueous solution
dc.subjectatomic absorption spectrometry
dc.subjectcalibration
dc.subjectcombustion
dc.subjectconcentration (parameters)
dc.subjectelution
dc.subjectflame atomic absorption spectrometry
dc.subjectnebulizer
dc.subjectonline system
dc.subjectpH
dc.subjectsensitivity and specificity
dc.subjectwater sampling
dc.subjectCadmium
dc.subjectCalibration
dc.subjectHydrochloric Acid
dc.subjectHydrogen-Ion Concentration
dc.subjectIons
dc.subjectModels, Chemical
dc.subjectSilicon Dioxide
dc.subjectSpectrophotometry
dc.subjectSpectrophotometry, Atomic
dc.subjectSpectrophotometry, Infrared
dc.subjectThiazoles
dc.subjectTime Factors
dc.subjectWater
dc.titleDetermination of cadmium in river water samples by flame AAS after on-line preconcentration in mini-column packed with 2-aminothiazole-modified silica gelen
dc.typeArtigo
dcterms.licensehttp://www.jsac.or.jp/analsci/info.php/about.html
dspace.entity.typePublication
unesp.author.lattes6981448637456391[6]
unesp.author.orcid0000-0003-4179-0574[8]
unesp.author.orcid0000-0003-4179-0574[6]
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Biociências, Botucatupt
unesp.campusUniversidade Estadual Paulista (UNESP), Instituto de Química, Araraquarapt
unesp.departmentFísica e Biofísica - IBBpt
unesp.departmentQuímica e Bioquímica - IBBpt
unesp.departmentQuímica Analítica - IQARpt

Arquivos

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
2-s2.0-3542992106.pdf
Tamanho:
104.9 KB
Formato:
Adobe Portable Document Format