Controle de qualidade e dosimetria em equipamentos de tomografia computadorizada

dc.contributor.authorPina, Diana Rodrigues de [UNESP]
dc.contributor.authorDuarte, Sérgio Barbosa
dc.contributor.authorNetto, Thomaz Ghilardi
dc.contributor.authorMorceli, José [UNESP]
dc.contributor.authorCarbi, Eros Duarte Ortigoso [UNESP]
dc.contributor.authorSouza, Rafael Toledo Fernandes de [UNESP]
dc.contributor.authorCosta Neto, André [UNESP]
dc.contributor.authorRibeiro, Sérgio Marrone [UNESP]
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.contributor.institutionCentro Brasileiro e Pesquisas Físicas
dc.contributor.institutionUniversidade de São Paulo (USP)
dc.date.accessioned2014-05-20T13:33:34Z
dc.date.available2014-05-20T13:33:34Z
dc.date.issued2009-06-01
dc.description.abstractOBJETIVO: Avaliação de condições dos equipamentos e dosimetria em setores de tomografia computadorizada utilizando protocolos de cabeça, abdome e coluna lombar em pacientes adultos (em três equipamentos distintos) e pediátricos com até um ano e meio de vida (em um dos equipamentos avaliados). MATERIAIS E MÉTODOS: Foram estimados o índice de dose em tomografia computadorizada e a dose média em cortes múltiplos, em exames com pacientes adultos, em três distintos equipamentos. Ainda foram estimadas as doses na superfície de entrada e as doses absorvidas em exame de cabeça para pacientes adultos e pediátricos em um dos equipamentos avaliados. RESULTADOS: Foram realizados testes de controle de qualidade, mecânicos, demonstrando que os equipamentos satisfazem as especificações de uso estabelecidas pelas normas vigentes. Os resultados da dosimetria mostraram que valores de dose média em cortes múltiplos excederam em até 109,0% os valores de níveis de referência, apresentando consideráveis variações entre os equipamentos avaliados neste estudo. As doses absorvidas obtidas com protocolos pediátricos são inferiores aos de pacientes adultos, apresentando redução de até 51,0% na tireoide. CONCLUSÃO: Neste estudo foram avaliadas as condições de operação de três equipamentos tomográficos, estabelecendo quais parâmetros devem ser trabalhados para a implantação de um programa de controle de qualidade nas instituições onde esta pesquisa foi desenvolvida.pt
dc.description.abstractOBJECTIVE: Evaluation of equipment conditions and dosimetry in computed tomography services utilizing protocols for head, abdomen, and lumbar spine in adult patients (in three different units) and pediatric patients up to 18 months of age (in one of the units evaluated). MATERIALS and METHODS: Computed tomography dose index and multiple-scan average dose were estimated in studies of adult patients with three different units. Additionally, entrance surface doses as well as absorbed dose were estimated in head studies for both adult and pediatric patients in a single computed tomography unit. RESULTS: Mechanical quality control tests were performed, demonstrating that computed tomography units comply with the equipment-use specifications established by the current standards. Dosimetry results have demonstrated that the multiple-scan average dose values were in excess of up to 109.0% the reference levels, presenting considerable variation amongst the computed tomography units evaluated in the present study. Absorbed doses obtained with pediatric protocols are lower than those with adult patients, presenting a reduction of up to 51.0% in the thyroid gland. CONCLUSION: The present study has analyzed the operational conditions of three computed tomography units, establishing which parameters should be set for the deployment of a quality control program in the institutions where this study was developed.en
dc.description.affiliationUniversidade Estadual Paulista Faculdade de Medicina de Botucatu
dc.description.affiliationCentro Brasileiro e Pesquisas Físicas
dc.description.affiliationUniversidade de São Paulo Faculdade de Medicina de Ribeirão Preto Hospital das Clínicas
dc.description.affiliationUniversidade Estadual Paulista Faculdade de Medicina de Botucatu Hospital das Clínicas
dc.description.affiliationUniversidade Estadual Paulista Instituto de Biociências de Botucatu
dc.description.affiliationUnespUniversidade Estadual Paulista Faculdade de Medicina de Botucatu
dc.description.affiliationUnespUniversidade Estadual Paulista Faculdade de Medicina de Botucatu Hospital das Clínicas
dc.description.affiliationUnespUniversidade Estadual Paulista Instituto de Biociências de Botucatu
dc.format.extent171-177
dc.identifierhttp://dx.doi.org/10.1590/S0100-39842009000300010
dc.identifier.citationRadiologia Brasileira. Colégio Brasileiro de Radiologia e Diagnóstico por Imagem, v. 42, n. 3, p. 171-177, 2009.
dc.identifier.doi10.1590/S0100-39842009000300010
dc.identifier.fileS0100-39842009000300010.pdf
dc.identifier.issn0100-3984
dc.identifier.scieloS0100-39842009000300010
dc.identifier.urihttp://hdl.handle.net/11449/11499
dc.language.isopor
dc.publisherColégio Brasileiro de Radiologia e Diagnóstico por Imagem
dc.relation.ispartofRadiologia Brasileira
dc.relation.ispartofsjr0,431
dc.rights.accessRightsAcesso aberto
dc.sourceSciELO
dc.subjectEquipamento tomográficopt
dc.subjectControle de qualidadept
dc.subjectDosimetriapt
dc.subjectRedução de dosept
dc.subjectComputed tomography equipmenten
dc.subjectQuality controlen
dc.subjectDosimetryen
dc.subjectDose reductionen
dc.titleControle de qualidade e dosimetria em equipamentos de tomografia computadorizadapt
dc.title.alternativeQuality control and dosimetry in computed tomography unitsen
dc.typeArtigo
dspace.entity.typePublication

Arquivos

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
S0100-39842009000300010.pdf
Tamanho:
217.2 KB
Formato:
Adobe Portable Document Format

Licença do Pacote

Agora exibindo 1 - 2 de 2
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: