Internal standard addition calibration: Determination of calcium and magnesium by atomic absorption spectrometry

dc.contributor.authorFortunato, Felipe Manfroi [UNESP]
dc.contributor.authorBechlin, Marcos A. [UNESP]
dc.contributor.authorNeto, José A. Gomes [UNESP]
dc.contributor.authorDonati, George L.
dc.contributor.authorJones, Bradley T.
dc.contributor.institutionUniversidade Estadual Paulista (Unesp)
dc.date.accessioned2018-12-11T16:38:27Z
dc.date.available2018-12-11T16:38:27Z
dc.date.issued2015-09-01
dc.description.abstractInternal standard addition (ISA) is a new calibration method that combines the principles of internal standardization and standard additions. The current work demonstrated the effectiveness of ISA for the determination of Ca and Mg in biodiesel samples and certified reference materials (CRMs) by flame atomic absorption spectrometry. Manganese and Sr were selected as internal standards for Mg and Ca, respectively. Results for Ca and Mg in CRMs using ISA were in agreement with certified values at the 95% confidence level (. t-test). The relative standard deviations (. n=. 12) were ca. 6% for both analytes. For comparison purposes, Ca and Mg were also determined by the traditional methods of external standard calibration (ES), standard additions (SA) and internal standardization (IS). Recoveries obtained with ISA (Ca: 93-109%; Mg: 100-106%) were similar to those found with IS (Ca: 100-112%; Mg: 98-105%), but significantly better than ES (Ca: 219-291%; Mg: 111-120%) and SA (Ca: 97-127%; Mg: 106-128%). Results for Ca and Mg determined in biodiesel and CRMs using ISA were more accurate and more precise than those obtained with ES, SA and IS.en
dc.description.affiliationSão Paulo State University-Unesp, Analytical Chemistry Department, P.O. Box 355
dc.description.affiliationDepartment of Chemistry, Wake Forest University
dc.description.affiliationUnespSão Paulo State University-Unesp, Analytical Chemistry Department, P.O. Box 355
dc.format.extent63-69
dc.identifierhttp://dx.doi.org/10.1016/j.microc.2015.04.009
dc.identifier.citationMicrochemical Journal, v. 122, p. 63-69.
dc.identifier.doi10.1016/j.microc.2015.04.009
dc.identifier.file2-s2.0-84928521423.pdf
dc.identifier.issn0026-265X
dc.identifier.scopus2-s2.0-84928521423
dc.identifier.urihttp://hdl.handle.net/11449/167816
dc.language.isoeng
dc.relation.ispartofMicrochemical Journal
dc.rights.accessRightsAcesso aberto
dc.sourceScopus
dc.subjectAccuracy
dc.subjectInternal standard addition calibration
dc.subjectPrecision
dc.titleInternal standard addition calibration: Determination of calcium and magnesium by atomic absorption spectrometryen
dc.typeArtigo