On-line monitoring of electrochemical degradation of paracetamol through a biomimetic sensor

Carregando...
Imagem de Miniatura

Data

2011

Autores

Oliveira, Mariana Calora Quintino de [UNESP]
Lanza, Marcos Roberto Vasconcelos
Jara, José Luis Paz
Sotomayor, Maria Del Pilar Taboada [UNESP]

Título da Revista

ISSN da Revista

Título de Volume

Editor

Resumo

This paper reports, for the first time, the online monitoring to the electrochemical degradation of the paracetamol using a biomimetic sensor coupled to a Flow Injection Analysis (FIA) system. The electrochemical degradation of the drug was carried out in aqueous medium using a flow-by reactor with a DSA anode. The process efficiency was monitored at real time by the biomimetic sensor constructed by modifying a glassy carbon electrode with a Nafion membrane doped with iron tetrapyridinoporphyrazine (FeTPyPz). Simultaneously, we carried out off-line analysis by liquid chromatography (HPLC) during the experiments in order to validate the proposed system. In addition, to investigate the degradation products of the paracetamol electrolysis, we used the techniques of UPLC/MS and GC/MS.

Descrição

Palavras-chave

Como citar

International Journal of Electrochemistry, v. 2011, p. 1-11, 2011.